Medical devices, healthcare technology and the business of care
510(k) clearance · K261959

Neptune SafeAir 10ft Smoke Evacuation Pencil, Square Coated, 70mm Blade, Push Button Switch (0703-046-100);Neptune SafeAir 10ft Smoke Evacuation Pencil, Square Coated, Push Button Switch, Bulk Non-Sterile (0703-046-102);Neptune SafeAir 10ft Smoke Evacuation Pencil, Square Coated, 70mm Blade, Rocker Switch (0703-047-100);Neptune SafeAir 10ft Smoke Evacuation Pencil, Square Coated, Rocker Switch, Bulk Non-Sterile (0703-047-102)

Stryker Instruments · Decision date Sep 17, 2026

MEDTECHLEDGER EXPLANATIONK261959

What did the FDA decide?

FDA recorded a substantially equivalent (SESE) decision on September 17, 2026 for K261959, submitted by Stryker Instruments of Portage, Michigan, under 21 CFR 878.4400 with product code GEI and the General & Plastic Surgery review panel. The subject is the Neptune® SafeAir® 10ft Smoke Evacuation Pencil, Square Coated, listed in four variants distinguished by 70mm blade presence, push button or rocker switch, and bulk non-sterile packaging. The summary describes single-use RF monopolar pencils for cutting and coagulation that also remove electrosurgery-generated smoke when paired with an effective evacuation system.

The listed predicate is Stryker's Neptune E-SEP Pencils (K160693). The summary states the subject adds a pencil variation with a coated electrode and squared edge tip, and that verification and validation results show these differences raise no new safety or effectiveness questions. Non-clinical testing listed includes sterilization, functional/performance, biocompatibility, thermal spread, packaging, and electrical safety and EMC. The source does not disclose clinical outcome data, the specific predicate-to-subject test results, or the full indications enclosure text; SESE is a clearance, not a PMA approval.

Regulatory pathway
510(k) premarket notification
Clinical specialty
General, Plastic Surgery
Decision date
Sep 17, 2026

FDA source record · MedtechLedger’s explanation describes the recorded change.

Regulatory record facts

Identifier
K261959
Regulatory pathway
510(k) premarket notification
Product code
GEI
Decision code
SESE
Decision date
Sep 17, 2026
Source update
Nov 9, 2026
Original record
Open FDA source record

About this medical device

FDA device category: Electrosurgical, Cutting & Coagulation & Accessories.

Product code
GEI
FDA device class
2

FDA decision timeline

  1. K261959 received

    Application receipt date recorded by FDA.

  2. K261959 · FDA decision

    SESE

Related devices and submissions

Explore the linked device profile

Other records with product code GEI

A shared product code identifies a classification relationship; it does not establish clinical equivalence.

Regulatory context

A 510(k) submission seeks a finding of substantial equivalence to a legally marketed device. A clearance and a PMA approval are different regulatory decisions.

Read our guide to FDA clearance and approval · FDA’s explanation of this pathway

Manufacturer intelligence

Stryker Instruments

Headquarters
Not reported
Brand website
stryker.com
FDA applicant address
Portage, MI

Sources and methodology

FDA fields and approval-order wording are presented as source information. MedtechLedger explanations are identified separately and describe the scope of the recorded decision. Indications from an original PMA are historical unless confirmed against subsequent labeling and supplements.

How we use FDA records