What did the FDA decide?

The FDA recorded a substantially equivalent (SESE) decision on October 29, 2026 for the Symmetrix™ PX Spinal System, a traditional 510(k) submitted by Medtronic Sofamor Danek USA, Inc. under K262572. The device is an intervertebral body fusion device (product code MAX, 21 CFR 888.3080, Class II, orthopedic panel) made of titanium alloy, cobalt chromium alloy, and PEEK, offered in varied widths, lengths, heights and lordotic angles, with optional Titan Surface Technology textures and compatibility with Medtronic Stealth™ navigation. SE denotes clearance to market, not PMA approval.

The submission lists the Catalyft™ PL Expandable Interbody System as primary predicate, with Catalyft™ LS, Adaptix, and Globus Rise Spacer as additional predicates. The summary reports bench testing (ASTM F2077, F2267, F2503) meeting predicate-derived acceptance criteria and MR Conditional labeling. The indications enclosure is referenced but not reproduced in the available text, so the cleared indications wording is not verified here; the summary's stated lumbar L2-S1 indications should be checked against that enclosure.

Regulatory pathway
510(k) premarket notification
Clinical specialty
Orthopedic
Decision date
Oct 29, 2026

FDA source record · MedtechLedger’s explanation describes the recorded change.

Regulatory record facts

Identifier
K262572
Regulatory pathway
510(k) premarket notification
Product code
MAX
Decision code
SESE
Decision date
Oct 29, 2026
Source update
Nov 9, 2026
Original record
Open FDA source record

About this medical device

FDA device category: Intervertebral Fusion Device With Bone Graft, Lumbar.

Product code
MAX
FDA device class
2

FDA decision timeline

  1. K262572 received

    Application receipt date recorded by FDA.

  2. K262572 · FDA decision

    SESE

Related devices and submissions

Explore the linked device profile

Other records with product code MAX

A shared product code identifies a classification relationship; it does not establish clinical equivalence.

Regulatory context

A 510(k) submission seeks a finding of substantial equivalence to a legally marketed device. A clearance and a PMA approval are different regulatory decisions.

Read our guide to FDA clearance and approval · FDA’s explanation of this pathway

Manufacturer intelligence

Medtronic Sofamor DANEK USA, Inc.