Precision Spinal Cord Stimulator (SCS) Systems
Boston Scientific Corp · Decision date Sep 19, 2025
What did the FDA decide?
FDA records P030017/S386 as a supplement to PMA P030017, with the supplement type “30-Day Notice”. The recorded decision is OK30 (30-day notice), dated Sep 19, 2025.
FDA approval order statement
the qualification of an alternate supplier, AVNA, for the IPG Pocket Template used with the Boston Scientific Neuromodulation Spinal Cord Stimulator and Deep Brain Stimulation Implantable Pulse Generators (IPGs). Additionally, the alternate supplier is changing the manufacturing process for the IPG Pocket Templates from chemical etching to metal stamping and laser etching
- Regulatory pathway
- PMA supplement
- Supplement type
- 30-Day Notice
- Clinical specialty
- Neurology
- Decision date
- Sep 19, 2025
- FDA change category
- Process Change - Manufacturer/Sterilizer/Packager/Supplier
Regulatory record facts
- Identifier
- P030017/S386
- Regulatory pathway
- PMA supplement
- Applicant
- Boston Scientific Corp
- Product code
- LGW
- Decision code
- OK30
- Decision date
- Sep 19, 2025
- Source update
- Nov 9, 2026
- Original record
- Open FDA source record
About this medical device
FDA device category: Stimulator, spinal-cord, totally implanted for pain relief.
- Applicant
- Boston Scientific Corp
- Product code
- LGW
- FDA device class
- 3
For the original device and its historical approval scope, see PMA P030017. Subsequent supplements may change the device or its indications.
FDA decision timeline
P030017 · Original PMA
PRECISION SPINAL CORD STIMULATION(SCS) SYSTEM
P030017/S386 received
Application receipt date recorded by FDA.
P030017/S386 · FDA decision
30-Day Notice
P030017/S389 · Related supplement
WaveWriter Alpha Prime Spinal Cord Stimulation (SCS) Systems
P030017/S390 · Related supplement
Precision Spinal Cord Stimulator (SCS) Systems
P030017/S374 · Related supplement
WaveWriter Alpha Spinal Cord Stimulator (SCS) Mixed Systems, M8 Adaptor
P030017/S377 · Related supplement
Precision Spinal Cord Stimulator (SCS) Systems
P030017/S391 · Related supplement
Precision Spinal Cord Stimulator (SCS) Systems
P030017/S387 · Related supplement
Precision Spinal Cord Stimulator (SCS) Systems Introducer Hub Mold Change
P030017/S394 · Related supplement
Precision Spinal Cord Stimulator (SCS) Systems
P030017/S395 · Related supplement
Precision Spinal Cord Stimulator (SCS) Systems
P030017/S392 · Related supplement
Celestia Implantable Pulse Generator, Celestia External Trial Stimulator
P030017/S393 · Related supplement
Spinal Cord Stimulation (SCS)
P030017/S397 · Related supplement
Precision Spinal Cord Stimulator (SCS) Systems
P030017/S398 · Related supplement
Precision Spinal Cord Stimulator (SCS) Systems
P030017/S396 · Related supplement
WaveWriter Alpha and WaveWriter Alpha Prime SCS Systems; Celestia and Celestia Prime SCS Systems
P030017/S399 · Related supplement
Precision Spectra SCS Systems; Precision Novi SCS Systems; Precision Montage MRI SCS Systems; Spectra Wave Writer SCS Sy
Timeline entries are records held by MedtechLedger for P030017. View FDA’s full supplement list.
Related devices and submissions
Same PMA: P030017
Explore the linked device profile
Other records with product code LGW
| Product / regulatory record | Company | Pathway | Decision date | Source |
|---|---|---|---|---|
| ABBOTT MEDICAL | PMA supplement | Oct 30, 2026 | FDA | |
| ABBOTT MEDICAL | PMA supplement | Oct 30, 2026 | FDA | |
| Medtronic Neuromodulation | PMA supplement | Oct 30, 2026 | FDA | |
| Medtronic Neuromodulation | PMA supplement | Oct 29, 2026 | FDA | |
| ABBOTT MEDICAL | PMA supplement | Oct 28, 2026 | FDA |
A shared product code identifies a classification relationship; it does not establish clinical equivalence.
Regulatory context
A PMA supplement records a change to a device with an existing premarket approval. The supplement’s approval applies to the change described in that submission.
Read our guide to FDA clearance and approval · FDA’s explanation of this pathway
Manufacturer intelligence
Boston Scientific Corp
- Headquarters
- Not reported
- Brand website
- bostonscientific.com
- FDA applicant address
- 25155 Rye Canyon Loop, Valencia, CA, 91355