Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0033-2026

Medtronic, Simplera Sensor, REF: MMT-5100JD1 Recall

Medtronic, Simplera Sensor, REF: MMT-5100JD1

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medtronic, Simplera Sensor, REF: MMT-5100JD1. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0033-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 7, 2025
FDA report date
Oct 8, 2025
Quantity in commerce
1,440 units
Distribution
See key details
FDA cause
Process design

What is being recalled?

Medtronic, Simplera Sensor, REF: MMT-5100JD1

Affected model & UDI checker

Search the exact identifiers published for Z-0033-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MMT-5100JD1Medtronic, Simplera Sensor, REF: MMT-5100JD1Check source noticeNot specified
Not specifiedMedtronic, Simplera Sensor, REF: MMT-5100JD1Check source notice20763000649580
FDA code information

Lot # HG81GLZ/ GTIN: 20763000649580

View the original FDA recall notice

Why was this product recalled?

The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection.

What actions did Medtronic MiniMed, Inc. take?

On May 7, 2025, Medtronic issued a "Urgent Field Safety Notice" to affected consignees. Medtronic asked consignees to take the following actions: 1. Verify if the sensors you currently have are from the impacted lot. 2. If you have a Simplera" sensor from the impacted lot, do not use this Simplera" sensor and dispose of it according to local regulations. 3. Please contact our helpline/your Medtronic contact at 01923 205167 to request a replacement. 4. Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter, even if your sensor is unaffected, or if you do not have any sensors from the affected lot.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology