Brand Name: Expressaire Product Name:DISPOSABLE BB TOURNQUET CUFF, 26" X 6" DUAL BLADDER, 1-LINE (CS 12) ORANGE Recall

Brand Name: Expressaire Product Name:DISPOSABLE BB TOURNQUET CUFF, 26" X 6" DUAL BLADDER, 1-LINE (CS 12) ORANGE Model/Catalog Number: A70107016 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Expressaire Product Name:DISPOSABLE BB TOURNQUET CUFF, 26" X 6" DUAL BLADDER, 1-LINE (CS 12) ORANGE. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0036-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 24, 2026
FDA report date
Oct 7, 2026
Quantity in commerce
36
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: Expressaire Product Name:DISPOSABLE BB TOURNQUET CUFF, 26" X 6" DUAL BLADDER, 1-LINE (CS 12) ORANGE Model/Catalog Number: A70107016 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-0036-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
A70107016Brand Name: Expressaire Product Name:DISPOSABLE BB TOURNQUET CUFF, 26" X 6" DUAL BLADDER, 1-LINE (CS 12) ORANGECheck source noticeNot specified
FDA code information

Lot Code: Model Number: A70107016 UDI-DI: (01) 00885914035603 All Lots produced from 01SEP2016 to 01MAR2023: P0325STY

View the original FDA recall notice

Why was this product recalled?

Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.

What actions did ASP Global, LLC. take?

On 08/24/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was emailed to customers informing them that, ASP Global identified a discrepancy between the labeled shelf life and available shelf-life validation data for certain Sterile Disposable Tourniquet Cuff products. While affected products were labeled with a 10-year shelf life, available validation documentation supports a shelf life of 5-years. Customer are instructed to: a) Review Appendix A to identify affected item codes and lot numbers. b) Discontinue use of the affected products. c) Quarantine any remaining impacted inventory to prevent further use. d) Complete and return the enclosed Recall Response Form to confirm receipt and actions to [email protected]. e) Dispose remaining inventory in accordance with the local facility s guidelines or notify ASP Global at [email protected] for return authorization and disposal assistance. f) If customers have distributed affected products to other facilities or customers, notify those recipients of this recall and provide them with a copy of this notification. For questions regarding this recall, contact: ASP Global Field Actions Team Email: [email protected] Phone: (855)-867-2190| Mon Fri, 9:00 AM 5:00 PM EST

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology