ClariTEE Miniaturized TEE probe. Recall

ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.

FDA recall Class II

Ongoing

FDA lists this Class II recall for ClariTEE Miniaturized TEE probe.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0038-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 29, 2025
FDA report date
Oct 15, 2025
Quantity in commerce
45 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.

Affected model & UDI checker

Search the exact identifiers published for Z-0038-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CLT-010ClariTEE Miniaturized TEE probe.Check source noticeNot specified
Not specifiedClariTEE Miniaturized TEE probe.Check source notice00861589000108
FDA code information

Model CLT-010; UDI-DI 00861589000108; Lot number: 20261209.

View the original FDA recall notice

Why was this product recalled?

Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did ImaCor Inc. take?

ImaCor, Inc. notified consignees on 08/29/2025 via email. Consignees were instructed to Immediately discontinue use of the affected ClariTEE probes, quarantine all affected units in inventory, confirm receipt of this notice, return all affected units to ImaCor, and complete and return the response form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology