EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE P180 Recall
EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE P180
FDA recall Class I
Ongoing
FDA lists this Class I recall for EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE P180. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0056-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Sep 11, 2025
- FDA report date
- Oct 22, 2025
- Quantity in commerce
- 1,070 units US; 2,180 units OUS
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE P180
Affected model & UDI checker
Search the exact identifiers published for Z-0056-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| BF-P180 | EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE P180 | Check source notice | Not specified |
| Not specified | EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE P180 | Check source notice | 4953170339288 |
FDA code information
Model No. BF-P180; UDI: 4953170339288; All Serial No.
Why was this product recalled?
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
What actions did Olympus Corporation of the Americas take?
An URGENT Medical Device Correction notification dated 9/11/25 was mailed to consignees to notify them of this recall action. Consignees are asked to carefully read the recall notification and identify any affected devices in their inventory. Consignees are to ensure all personnel are knowledgeable and aware of the updated IFU. Consignees are to acknowledge receipt of the recall notification through Olympus' web portal at https://olympusamerica.com/recall. Forward the provided recall notification if product was further distributed. Consignees with any questions can contact Cynthia Ow by phone at 647-999-3203 or by email at [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0056-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology