Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0067-2027

Brand Name: Ferndale Laboratories; Medline; Custom Interventional Pain Management CIPM Product Name: Mastisol Skin Adhesive Recall

Brand Name: Ferndale Laboratories; Medline; Custom Interventional Pain Management CIPM Product Name: Mastisol Skin Adhesive Model/Catalog Number: 24161B (Ferndale and Medline); Tray 17025 17g continuous PBN tray (CIPM) Software Version: n/a Product Description: Skin Adhesive in a narrow plastic sterilized tube. Mastisol is a skin glue often used during surgical procedures. Continuous epidural trays are designed for use by anesthesiologists and radiologists in the administration of regional anesthesia. Component: surgical kits

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Ferndale Laboratories; Medline; Custom Interventional Pain Management CIPM Product Name: Mastisol Skin Adhesive. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0067-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 15, 2026
FDA report date
Oct 7, 2026
Quantity in commerce
56
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: Ferndale Laboratories; Medline; Custom Interventional Pain Management CIPM Product Name: Mastisol Skin Adhesive Model/Catalog Number: 24161B (Ferndale and Medline); Tray 17025 17g continuous PBN tray (CIPM) Software Version: n/a Product Description: Skin Adhesive in a narrow plastic sterilized tube. Mastisol is a skin glue often used during surgical procedures. Continuous epidural trays are designed for use by anesthesiologists and radiologists in the administration of regional anesthesia. Component: surgical kits

Affected model & UDI checker

Search the exact identifiers published for Z-0067-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
24161BBrand Name: Ferndale Laboratories; Medline; Custom Interventional Pain Management CIPM Product Name: Mastisol Skin AdhesiveCheck source noticeNot specified
FDA code information

Lot Code: CIPM product / reorder #17025; Description: 17g continuous PBN tray; Item #110026; Received Lot# 27022; Affected Lot #1225271; Expiration Date: 01/31/2028; DHR or Batch Record #64591; UDI: *+B87517LIDO/$$20131281225271T* UDI- DI #: B87517LID0 with Lidocaine

View the original FDA recall notice

Why was this product recalled?

Due to manufacturing issues with the liquid skin adhesive containing in surgical convenience trays/kits.

What actions did Custom Interventional Pain Management, LLC take?

On 01/15/2026, a customer notification letter was email to a single customer informing them that, the MASTISOL Liquid adhesive shipped in kits in lot 1225271, P.O.#2513972 was exposed to a double sterilization cycle by Ferndale Laboratories, Inc. which may contribute to the butyrate tubing being brittle and cracking during actuation. The customer was instructed to: 1. lmmediately examine your inventory and quarantine Lot 1225271. 2. Return all units of MASTISOL in Lot 1225271 to: CIPM 12460 Network Blvd. Suite 106 San Antonio, TX 78249 3. Complete the enclosed Return Materials Authorization Form QF-85-05-02 and return to [email protected] or [email protected]. For questions - call Toll Free 1-888-392-4514

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.