Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0069-2027

Generic Name: Kit, Urine, Clean Catch Trade Name: Kit: Urinary Clean Catch 20/cs Recall

Generic Name: Kit, Urine, Clean Catch Trade Name: Kit: Urinary Clean Catch 20/cs Model/Catalog Number: 76769 Lot Number: 333606 UDI-DI: 20809160118228 Software Version: N/A

FDA recall Class III

Ongoing

FDA lists this Class III recall for Generic Name: Kit, Urine, Clean Catch Trade Name: Kit: Urinary Clean Catch 20/cs. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0069-2027
Recall classification
Class III
Enforcement status
Ongoing
Initiation date
Jul 9, 2026
FDA report date
Oct 14, 2026
Quantity in commerce
4000
Distribution
See key details
FDA cause
Labeling design

What is being recalled?

Generic Name: Kit, Urine, Clean Catch Trade Name: Kit: Urinary Clean Catch 20/cs Model/Catalog Number: 76769 Lot Number: 333606 UDI-DI: 20809160118228 Software Version: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-0069-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
76769Generic Name: Kit, Urine, Clean Catch Trade Name: Kit: Urinary Clean Catch 20/csCheck source noticeNot specified
Not specifiedGeneric Name: Kit, Urine, Clean Catch Trade Name: Kit: Urinary Clean Catch 20/csCheck source notice20809160118228
FDA code information

Lot Code: Lot Number: 333606 UDI: 20809160118228 Expiration Date: 3/31/2027

View the original FDA recall notice

Why was this product recalled?

An internal data-entry error caused kit lot 333606 to be labeled with an expiration date of March 31, 2027, instead of February 28, 2027.

What actions did Medical Action Industries, Inc. 306 take?

On 07/09/2026, an "URGENT: MEDICAL DEVICE FIELD CORRECTION" Letter was emailed to customers informing them that, due to an internal error, item 76769, KIT: Urinary Clean Catch, was inadvertently assigned an incorrect expiration date of 2027/03/31. The correct expiration date is 2027/02/28. Product was distributed between 09/24/2025 to 12/17/2025. On 08/03/2026, an updated "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter was emailed to customer to notify them that, The previous notifications, dated July 9, 2026 and July 24, 2026, contained an incorrect lot number. As stated below, the correct lot number is 76769. Customers were instructed to: 1. Identify and quarantine all impacted product. 2. Add warning labels to impacted product with the new expiration date. A labeling template is enclosed for printing labels which can be affixed on all affected product in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. - Customers (end-users): If you received product from a distributor, contact the distributor or your sales representative with any questions you may have. - Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 3. Complete the attached Recall Response Form and return by August 21, 2026. For questions - contact Regulator Affairs at 844-425-9273 or email [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.