Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0078-2027

Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.); Recall

Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);

FDA recall Class II

Ongoing

FDA lists this Class II recall for Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0078-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 21, 2026
FDA report date
Oct 14, 2026
Quantity in commerce
539
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);

Affected model & UDI checker

Search the exact identifiers published for Z-0078-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
728333Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);Check source noticeNot specified
728340Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);Check source noticeNot specified
728344Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);Check source noticeNot specified
728345Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);Check source noticeNot specified
Not specifiedSpectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);Check source notice00884838111103
Not specifiedSpectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);Check source notice00884838117822
Not specifiedSpectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);Check source notice00884838123748
FDA code information

1. Model Number: 728333 (U.S.); UDI-DI: 1. 00884838101111, 2. 00884838127876; Serial Numbers: 12055, 12054, 12052, 398006, 12044, 12045, 10237, 10056, 10194, 12059, 10278, 10319, 10143, 10081, 12047, 12048, 10198, 12046, 12043, 10262, 12042, 12041, 10186, 10241, 12039, 10317, 10316, 10264, 10127, 12038, 12037, 396014, 12036, 398005, 10016, 12034, 10076, 10074, 12035, 10021, 10018, 10023, 12032, 12029, 398003, 398004, 396013, 12027, 12026, 12023, 12025, 12024, 12021, 12022, 12017, 12019, 12018, 12020, 10312, 10311, 10310, 10305, 12016, 10308, 10306, 10307, 12014, 12013, 10304, 10302, 12012, 12011, 12010, 396012, 12007, 10301, 12009, 10300, 10299, 12008, 12006, 12005, 10298, 12002, 12003, 10294, 10292, 12001, 12000, 10293, 10290, 396011, 396010, 10291, 10289, 10279, 10288, 10287, 10284, 10286, 10281, 10285, 10271, 10276, 396009, 10280, 10275, 10273, 10277, 396008, 396006, 396007, 10274, 10272, 10270, 10269, 10252, 10154, 10268, 10267, 10261, 10266, 10265, 10219, 10263, 10260, 10157, 10259, 10224, 10240, 10257, 10246, 10258, 10253, 10256, 10251, 10254, 10243, 10250, 10248, 10247, 10244, 10249, 10245, 10236, 10223, 10239, 10242, 10235, 10234, 10221, 10231, 10209, 10232, 10228, 10226, 10229, 10230, 10220, 10217, 10227, 10225, 10218, 10222, 10212, 10214, 10215, 10213, 10211, 10210, 396004, 10208, 10207, 10206, 10204, 10202, 10197, 10203, 10205, 10200, 10201, 10199, 10196, 10195, 10193, 10175, 10190, 10189, 10188, 10191, 10187, 10184, 10185, 10183, 10181, 10180, 10182, 10179, 10178, 10177, 10174, 10176, 10173, 10172, 10171, 10170, 10168, 10169, 10167, 10163, 10165, 10166, 10158, 10161, 10160, 10148, 10159, 10156, 10155, 10153, 10152, 10150, 10151, 10149, 10147, 10146, 10145, 10144, 10141, 10142, 10140, 10138, 10139, 10137, 10134, 10136, 10135, 10133, 10132, 10130, 10131, 10128, 10129, 10126, 10049, 10124, 10119, 10123, 10121, 10120, 10118, 10122, 10032, 10116, 10113, 10115, 10114, 10111, 10112, 10110, 10104, 396003, 396001, 10109, 10108, 10107, 10106, 10105, 10097, 10103, 10100, 10102, 10101, 10098, 10099, 10090, 10096, 10095, 10093, 10094, 10091, 10086, 10088, 10085, 10089, 10087, 10084, 10082, 10080, 10077, 10078, 10075, 10069, 10072, 10071, 10070, 10067, 10068, 10066, 10065, 10062, 10061, 10060, 10057, 10058, 10052, 10055, 10053, 10050, 10047, 10045, 10048, 10044, 10042, 10043, 10041, 10040, 10039, 10038, 10036, 10035, 10034, 10031, 10033, 10027, 10024, 10020, 10019, 10017, 10015, 10013, 10117, 10079, 10314, 12028, 12015, 10303, 10233, 10162, 12033, 12031, 12030, 10092, 10046, 10125, 10063, 10025; 2. Model Number: 728340 (Outside U.S.); UDI-DI: 00884838111103; Serial Numbers: 397035, 397004, 397168, 397167, 397164, 397159, 397160, 397155, 397153, 397152, 397150, 397151, 397148, 397147, 397146, 397145, 397141, 397142, 397139, 397140, 397137, 397133, 397132, 397130, 397126, 397125, 397123, 397121, 397120, 397117, 397116, 397115, 397112, 397110, 397108, 397107, 397106, 397102, 397101, 397099, 397098, 397096, 397095, 397093, 397091, 397085, 397089, 397088, 397087, 397086, 397084, 397083, 397082, 397080, 397081, 397079, 397077, 397076, 397075, 397073, 397072, 397071, 397070, 397067, 397066, 397065, 397062, 397064, 397063, 397060, 397061, 397059, 397058, 397056, 397055, 397054, 397053, 397050, 397049, 397047, 397048, 397046, 397045, 397043, 397044, 397041, 397040, 397039, 397038, 397037, 397036, 397034, 397033, 397032, 397026, 397031, 397029, 397028, 397027, 397025, 397024, 397023, 397022, 397021, 397020, 397018, 397019, 397016, 397017, 397015, 397014, 397012, 397009, 397013, 397011, 397003, 397010, 397008, 397006, 397002, 397005, 397007, 397105, 397030, 397128, 397119, 397097; 3. Model Number: 728344 (Outside U.S.); UDI-DI: 00884838117822; Serial Numbers: 397170, 397162, 397158, 397149, 397166, 397165, 397163, 397161, 397156, 397143, 397136, 397135, 397131, 397129, 397127, 397122, 397124, 397118, 397114, 397113, 397111, 397104, 397100, 397094, 397092, 397090, 397078, 397074, 397068, 397057, 397103; 4. Model Number: 728345 (Outside U.S.); UDI-DI: 00884838123748; Serial Numbers: 14004, 399003, 14001, 14000;

View the original FDA recall notice

Why was this product recalled?

Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did PHILIPS MEDICAL SYSTEMS take?

On August 21, 2026, URGENT Medical Device Correction letters were sent to customers. Customer actions: User Instructions Philips recommends the following actions be taken for continued safe use of the device: 1. Continue to use your system(s) in accordance with the intended use; 2. Circulate this notice to all users of the device so that they are aware of the product issues, associated potential hazards, and manufacturer recommendations (refer to Table 1 and Table 2 for specific details regarding recommended actions); 3. Post this notice near the affected unit(s) for ease of reference until a solution is installed on your system. 4. Please complete and return the attached response form within 30 days to confirm receipt and understanding of this notice.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology