Sperm Ab ELISA; Recall
Sperm Ab ELISA; Model/Catalog Number: EIA-1826; Product Description: Sperm Ab ELISA (96 well plate test);
FDA recall Class II
Ongoing
FDA lists this Class II recall for Sperm Ab ELISA;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0082-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 2, 2026
- FDA report date
- Oct 14, 2026
- Quantity in commerce
- 211
- Distribution
- See key details
- FDA cause
- Reprocessing Controls
What is being recalled?
Sperm Ab ELISA; Model/Catalog Number: EIA-1826; Product Description: Sperm Ab ELISA (96 well plate test);
Affected model & UDI checker
Search the exact identifiers published for Z-0082-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| EIA-1826 | Sperm Ab ELISA; | Check source notice | 04048474018264 |
FDA code information
Model/Catalog Number: EIA-1826; UDI: 04048474018264; Lot# 320K046;
Why was this product recalled?
Potential for false negative patient results.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did DRG International, Inc. take?
On July 2, 2026 Field Safety Notice letters were sent to customers. Actions to be taken: Please check your stocks for the lot in question (320K046), stop using these kits immediately and quarantine these products without delay. Please destroy the kits and send us the confirmation with the completed response form. If you have transferred any of the affected products to another laboratory, please provide them with a copy of this letter.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0082-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology