DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520 Recall
DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520
FDA recall Class II
Ongoing
FDA lists this Class II recall for DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0084-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 2, 2026
- FDA report date
- Oct 14, 2026
- Quantity in commerce
- 146 US, 170 OUS
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520
Affected model & UDI checker
Search the exact identifiers published for Z-0084-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520 | Check source notice | 14987666545058 |
| Not specified | DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520 | Check source notice | 14987666545065 |
FDA code information
Software version 1.4.5 or earlier. 1. UDI-DI 14987666545058 2. UDI-DI 14987666545065
Why was this product recalled?
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
What actions did BECKMAN COULTER, INC. take?
On September 2, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to inform their laboratory or medical director to assess laboratory and patient impact. For analyzers on software versions earlier than 1.5: - Users should verify that their instrument displays the intended units in the Test Menu page within the System Configuration page prior to reporting patient results. - Recall letter should be posted on or near affected analyzers if they have the affected software version. This issue is resolved in software version 1.5. If your analyzer is currently operating on an earlier software version, your Beckman Coulter service representative will contact you to schedule the upgrade to software version 1.5.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0084-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.