BARD EP XT DECAPOLAR STEERABLE, DIAGNOSTIC EP CATHETER, 6F, LARGE 4.0, 10, 2-5-2MM, REF: 201007 Recall
BARD EP XT DECAPOLAR STEERABLE, DIAGNOSTIC EP CATHETER, 6F, LARGE 4.0, 10, 2-5-2MM, REF: 201007
FDA recall Class II
Ongoing
FDA lists this Class II recall for BARD EP XT DECAPOLAR STEERABLE, DIAGNOSTIC EP CATHETER, 6F, LARGE 4.0, 10, 2-5-2MM, REF: 201007. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0085-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 18, 2026
- FDA report date
- Oct 14, 2026
- Quantity in commerce
- 10
- Distribution
- See key details
- FDA cause
- Packaging process control
What is being recalled?
BARD EP XT DECAPOLAR STEERABLE, DIAGNOSTIC EP CATHETER, 6F, LARGE 4.0, 10, 2-5-2MM, REF: 201007
Affected model & UDI checker
Search the exact identifiers published for Z-0085-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 201007 | BARD EP XT DECAPOLAR STEERABLE, DIAGNOSTIC EP CATHETER, 6F, LARGE 4.0, 10, 2-5-2MM, REF: 201007 | Check source notice | Not specified |
FDA code information
UDI-DI: 00885825003838. Lot: 5640704, Serial: 5097742, Distribution Date: May 26th 2026
Why was this product recalled?
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could lead to particulate embolization, infection, thrombus or ganuloma formation from particulate matter.
What actions did Stryker Sustainability Solutions take?
On 8/18/2026, the recall notice was mailed to the customer asking them to do the following: 1) Inform individuals within your organization who need to be aware of this action. 2) Quarantine and return affected product to the firm. 4) Complete and return the Business Reply Form via email to [email protected] Questions can be directed to the firm via email, [email protected], or via phone 800-624-4422.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0085-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.