Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station Recall
Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0089-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 8, 2026
- FDA report date
- Oct 14, 2026
- Quantity in commerce
- 808
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-0089-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| CN-6201 | Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station | Check source notice | Not specified |
FDA code information
Lot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40
Why was this product recalled?
Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Nihon Kohden America LLC take?
On 09/08/2026, an "URGENT: MEDICAL DEVICE CORRECTION" Letter was sent via USPS certified mail to customers informing that, Nihon Kohden Corporation announced the discovery of a vulnerability within a limited number of CNS-6201 models to Denial of Service (DoS) attacks. The identified vulnerability is classified as a NULL pointer dereference* (CWE-476). This type of software error may cause the system to crash when processing malformed or unexpected network input. Nihon Kohden has not received any complaints related to this vulnerability. If a CNS-6201 is affected by a DoS event, the device may stop its monitoring function due to a system crash, resulting in temporary loss of central monitoring functionality. Customers are instructed to: 1. Identify any affected CNS-6201 units in their inventory. 2. Complete the attached Medical Device Notification Return Response Form and email the completed Medical Device Notification Return Response Form back to NKA Technical Services at [email protected]. Issue Resolution a. Nihon Kohden Corporation advises that the customer upgrade to a successor product (e.g. CNS-2101). b. For customers who continue to operate CNS-6201 systems, Nihon Kohden reiterates the importance of implementing established cybersecurity controls that are part of the intended deployment architecture and long-standing cybersecurity guidance for these systems. These controls are not newly introduced mitigations but rather represent expected deployment requirements designed to support the secure operation of the system throughout its lifecycle. For questions - call Nihon Kohden America at 1-800-325-0283, Option 2 (M-F, 8am-5pm PDT) or email [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0089-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology