Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0117-2027

TufTex Over-the-Wire Embolectomy Catheter; Recall

TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;

FDA recall Class II

Ongoing

FDA lists this Class II recall for TufTex Over-the-Wire Embolectomy Catheter;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0117-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 28, 2026
FDA report date
Oct 14, 2026
Quantity in commerce
54
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

TufTex Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;

Affected model & UDI checker

Search the exact identifiers published for Z-0117-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
E1651-88TufTex Over-the-Wire Embolectomy Catheter;Check source noticeNot specified
Not specifiedTufTex Over-the-Wire Embolectomy Catheter;Check source notice00840663100866
FDA code information

Catalog Number: E1651-88; UDI-DI: 00840663100866; Lot Number: XOT00002181;

View the original FDA recall notice

Why was this product recalled?

Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.

What actions did LeMaitre Vascular, Inc. take?

On August 28, 2026 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: 1. Check your inventory against the list of lots in this letter. Immediately quarantine any recalled devices. 2. Complete the form at the end of this letter. Please note that you must return the form even if you have no devices in inventory. 3. LeMaitre customer service will provide an RMA number to coordinate the product return. 4. When a recalled device has been returned to LeMaitre Vascular, a replacement device will be provided. 5. If you have transferred devices to another facility, please forward a copy of this recall letter to them.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology