HexaVue Focus Integration System, a single, all-in-one system that handles operating room video routing, surgical video integration, and environmental control, Software Version: 2.0.18 Recall
HexaVue Focus Integration System, a single, all-in-one system that handles operating room video routing, surgical video integration, and environmental control, Software Version: 2.0.18
FDA recall Class II
Ongoing
FDA lists this Class II recall for HexaVue Focus Integration System, a single, all-in-one system that handles operating room video routing, surgical video integration, and environmental control, Software Version: 2.0.18. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0118-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 23, 2026
- FDA report date
- Oct 14, 2026
- Quantity in commerce
- 58 systems
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
HexaVue Focus Integration System, a single, all-in-one system that handles operating room video routing, surgical video integration, and environmental control, Software Version: 2.0.18
Affected model & UDI checker
Search the exact identifiers published for Z-0118-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | HexaVue Focus Integration System, a single, all-in-one system that handles operating room video routing, surgical video integration, and environmental control, Software Version: 2.0.18 | Check source notice | 00724995245047 |
FDA code information
Lot Code: UDI: 00724995245047; Serial Numbers: RL25293001 RL25301003 RL25301004 RL26007031 RL26013044 RL26013045 RL26028062 RL26040066 RL26054029 RL26062001 RL26063005 RL26063007 RL26063013 RL26063014 RL26063019 RL26063021 RL26063022 RL26063023 RL26069033 RL26069034 RL26072049 RL26072052 RL26072053 RL26072055 RL26072056 RL26072056 RL26077072 RL26077073 RL26077074 RL26083077 RL26084078 RL26084080 RL26084081 RL26085083 RL26085084 RL26085085 RL26085086 RL26085087 RL26085088 RL26085089 RL26089090 RL26089091 RL26089092 RL26089093 RL26089094 RL26089095 RL26089096 RL26089097 RL26098098 RL26104099 RL26105102 RL26133103 RL26133104 RL26138105 RL26139106 RL26140114 RL26187116 RL26243121
Why was this product recalled?
A software issue may result in unintended mismatch of patient information on manually printed images. Incorrect patient information appearing on printed images could create a documentation discrepancy or incorrect association of clinical images with a patient record.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did STERIS CORPORATION take?
An URGENT MEDICAL DEVICE RECALL notification letter dated 9/9/26 was sent to customers. User Action Users can continue to use their integration system(s) prior to completion of the correction, but should confirm that manually printed images are correctly labeled prior to clinical use, filing or scanning into a patient record, distribution or other reliance upon the printed output. Customers should consider reviewing any manually printed images generated using software version 2.0.18 if they have reason to believe the manual print workflow described in this notice may have occurred. Prior to your system(s) receiving the updated software, if you are utilizing the manual printing functionality and have questions regarding the issue described in this notice or interim mitigation measures, please contact STERIS ORI Support at [email protected], or at 800-548-4873. Users are reminded to report any adverse events via FDA s MedWatch Database or local competent authority should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0118-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology