Cyclopen Microinterventional Cyclodialysis System. Recall
Cyclopen Microinterventional Cyclodialysis System. Model Number: CP-300. Product Description: The CycloPen Microinterventional Cyclodialysis device is a manual surgical instrument used for the ab-interno construction or modification of a cyclodialysis, followed by delivery of viscous materials, bio-tissue/allograft, or other materials to reinforce and maintain the cleft.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Cyclopen Microinterventional Cyclodialysis System.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0134-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 6, 2025
- FDA report date
- Oct 22, 2025
- Quantity in commerce
- 1,262 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Cyclopen Microinterventional Cyclodialysis System. Model Number: CP-300. Product Description: The CycloPen Microinterventional Cyclodialysis device is a manual surgical instrument used for the ab-interno construction or modification of a cyclodialysis, followed by delivery of viscous materials, bio-tissue/allograft, or other materials to reinforce and maintain the cleft.
Affected model & UDI checker
Search the exact identifiers published for Z-0134-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| CP-300 | Cyclopen Microinterventional Cyclodialysis System. | Check source notice | Not specified |
FDA code information
1. Model Number CP-300; UDI: +B962CP3000/$$7351645/16D20250201/14D20260201/Q1_; Lot Number: 351645; Exp date 02/2026. 2. Model Number CP-300; UDI: +B962CP3000/$$7351726/16D20250501/14D20260501/Q11; Lot Number 351726; Exp date 05/2026. 3. Model Number CP-300; UDI: +B962CP3000/$$7351795/16D20250601/14D20260601/Q19; Lot Number 351795; Exp date 06/2026.
Why was this product recalled?
Potential for compromised integrity of the sterile packaging.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Iantrek Inc. take?
Iantrek notified consignees on 09/06/2025 via email. Consignees were instructed to immediately remove and quarantine all affected units to prevent their use, notify personnel or customers if affected units were forwarded to other facilities. Consignees may also alternatively provide Iantrek with contact information so that they can notify them on the customer's behalf. An Iantrek representative will be contacting each consignee immediately to help facilitate removal, return and replacement of affected units. They will arrange to have replacement product shipped to consignees at no cost.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0134-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology