Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0134-2027

JenaValve Trilogy Delivery Catheter, Size 23MM REF: TRILOGY-DC-S Recall

JenaValve Trilogy Delivery Catheter, Size 23MM REF: TRILOGY-DC-S

FDA recall Class II

Ongoing

FDA lists this Class II recall for JenaValve Trilogy Delivery Catheter, Size 23MM REF: TRILOGY-DC-S. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0134-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 31, 2026
FDA report date
Oct 21, 2026
Quantity in commerce
111 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

JenaValve Trilogy Delivery Catheter, Size 23MM REF: TRILOGY-DC-S

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

GTIN - 00850000898134/ Lot: DS-040826-003 DS-042326-003, DS-042726-006, DS-042826-005, DS-050426-007, DS-050626-002, DS-050626-006, DS-050626-009, DS-050726-005

View the original FDA recall notice

Why was this product recalled?

Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.

What actions did JenaValve Technology, Inc. take?

On the dates of August 1, 4, 6, and 19th, JenaValve Technology Inc. issued a Urgent: Medical Device Recall notification via email. JenaValve asked consignees to take the following actions: 1. Review the information contained in this letter and ensure that anyone in your facility who needs to be aware of this notification reads this letter carefully. 2. Examine your inventory immediately to determine if you have Trilogy Delivery Catheters from the affected lots and quarantine the subject products. 3. JenaValve field team members will schedule a visit to retrieve the subject Trilogy Delivery Catheters. If you prefer to return the products yourself, please contact us at 866-588-8287 and we will provide you with an RGA number prior to you shipping them to: JenaValve Technology Inc., 4 Cromwell, Irvine, California 92618, USA; Attn: Quality Department; Reference: FA26001 4. Please complete and sign the Business reply form (BRF) of this letter and email it to [email protected] or give it to your JenaValve field representative. 5. Upon receipt of the completed BRF and returned product, JenaValve will provide replacement products. Incomplete BRFs cannot be processed. 6. If any of the subject Trilogy Delivery Catheters have been transferred to another facility, identify the receiving facility on the Business Reply Form, contact [email protected], and provide a copy of this letter to the relevant personnel.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology