FDA RECALL RECORD · Z-0141-2027

Brand Name: IOMAX Product Name: IOMAX Cortical Module Recall

Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: IOMAX Product Name: IOMAX Cortical Module. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0141-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 18, 2026
FDA report date
Oct 21, 2026
Quantity in commerce
23
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system

Affected model & UDI checker

Search the exact identifiers published for Z-0141-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
190296-200Brand Name: IOMAX Product Name: IOMAX Cortical ModuleCheck source noticeNot specified
FDA code information

Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Lot Code: These devices are identified by the serial numbers listed below: 19029607AA0326007 19029607AA0226013 19029607AA0426028 19029607AA0226010 19029607AA0226007 19029607AA0326026 19029607AA0226015 19029607AA0326020 19029607AA0326018 19029607AA0326021 19029607AA0326022 19029607AA0226009 19029607AA0526022 19029607AA0326017 19029607AA0326027 19029607AA0326016 19029607AA0226016 19029607AA0326028 19029607AA0326024 19029607AA0326025 19029607AA0326023 19029607AA0326031 19029607AA0326006

View the original FDA recall notice

Why was this product recalled?

Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.

What actions did Cadwell Industries Inc take?

On 08/18/2026, an "URGENT: MEDICAL DEVICE RECALL" Letter was emailed informing customers that, a manufacturing defect was identified in a limited number of IOMAX Cortical Modules. Following a recent change in printed circuit board assembly (PCBA) vendor, the defect was discovered during routine production processes on a limited number of PCBA s that were used in IOMAX Cortical Modules. The identified risk is specific to the Cortical Module and Cascade Surgical Studio (CSS). However, because Cortical Modules can be used with Zenith, Arc, and CSS, a module currently in use with Zenith or Arc could later be connected to CSS. Cadwell is therefore recalling all potentially affected Cortical Modules regardless of current host system. If the defective PCBA is present, it could potential cause unintentionally deliver electrical stimulation to the patient, or allow a small amount of unintended electrical current to reach the patient. Customers are instructed to return all identified devices to Cadwell Industries, Inc. Cadwell Industries, Inc. is voluntarily recalling and replacing or repairing all affected devices from the field.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology