Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0143-2027

NaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030; Recall

NaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;

FDA recall Class II

Ongoing

FDA lists this Class II recall for NaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0143-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 10, 2026
FDA report date
Oct 21, 2026
Quantity in commerce
403
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

NaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Material Number (UPN): M00535030; UDI-DI: 08714729856191; Lot Number: 38904309, 39017263, 39365322;

View the original FDA recall notice

Why was this product recalled?

Device was incorrectly manufactured with a guidewire exit port (referred to in the IFU as the RX port), which is not a design feature associated with this device configuration.

What actions did BOSTON SCIENTIFIC CORPORATION take?

On September 10, 2026, Urgent Medical Device Removal letters were sent to customers. Actions to be taken: 1. Immediately stop further use or distribution and segregate affected product (see Attachment 1 for identification of the affected devices). 2. Continue to follow the standard of care when monitoring patients treated with the product. 3. Complete and return the enclosed Reply Verification Tracking Form per the included instructions. 4. Return product to Boston Scientific in accordance with the enclosed instructions. 97657220-FA Advanix Pancreatic LW Pusher. 5. Forward this notice to relevant personnel and any facilities to which the product was transferred.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology