LOGIQ P10 series with software version R4.5.7 Model Number 5877534 Recall
LOGIQ P10 series with software version R4.5.7 Model Number 5877534
FDA recall Class II
Ongoing
FDA lists this Class II recall for LOGIQ P10 series with software version R4.5.7 Model Number 5877534. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0155-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 18, 2025
- FDA report date
- Oct 22, 2025
- Quantity in commerce
- 71 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
LOGIQ P10 series with software version R4.5.7 Model Number 5877534
Affected model & UDI checker
Search the exact identifiers published for Z-0155-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 5877534 | LOGIQ P10 series with software version R4.5.7 Model Number 5877534 | Check source notice | Not specified |
| Not specified | LOGIQ P10 series with software version R4.5.7 Model Number 5877534 | Check source notice | 00195278464644 |
FDA code information
Model 5877534 UDI-DI: 00195278464644, Serial Numbers: LPX441358 LPX441335 LPX440894 LPX440598 LPX440898 LPX441357 LPX441328 LPX441402 LPX441407 LPX441405 LPX441326 LPX441126 LPX441406 LPX441403 LPX441404 LPX441439 LPX441440 LPX441327 LPX441043 LPX441399 LPX441370 LPX441422 LPX441369 LPX441001 LPX441447 LPX440990 LPX441374 LPX441376 LPX441428 LPX441453 LPX441427 LPX440999 LPX441373 LPX441420 LPX441384 LPX441375 LPX441421 LPX441424 LPX441371 LPX441425 LPX441426 LPX441367 LPX441368 LPX441377 LPX441382 LPX441455 LPX441423 LPX441430 LPX441018 LPX441417 LPX441437 LPX441445 LPX441265 LPX441446 LPX441412 LPX441451 LPX441462 LPX441448 LPX441450 LPX441475 LPX441457 LPX441444 LPX490014 LPX441099 LPX441452
Why was this product recalled?
The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did GE Medical Systems, LLC take?
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 9/18/25 was sent to customers. Actions to be taken by Customer/User For LOGIQ P9 and LOGIQ P10 series ultrasound systems with system software version R4.5.7, discontinue using the UGAP feature (please see Appendix for product identification). You may continue using other imaging features. After the correction has been implemented, please destroy the installation media for software R4.5.7 at your site. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to [email protected]. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0155-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology