FDA RECALL RECORD · Z-0158-2026

Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723 Recall

Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723

FDA recall Class II

Ongoing

FDA lists this Class II recall for Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0158-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 18, 2025
FDA report date
Oct 22, 2025
Quantity in commerce
51,144
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723

Affected model & UDI checker

Search the exact identifiers published for Z-0158-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PT101USAirvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723Check source noticeNot specified
PT100USAirvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723Check source noticeNot specified
UI-185043723Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723Check source noticeNot specified
FDA code information

REF/UDI-DI: PT101US/9420012422347; PT100US/9420012422248. Disinfection Kit User Manual all revisions before UI-185043723 rev P

View the original FDA recall notice

Why was this product recalled?

When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.

What actions did Fisher & Paykel Healthcare, Ltd. take?

On 9/18/2025, correction notices were communicated via email and phone call. Customers were asked to do the following: 1) Prior to use on a new patient, follow the instructions of the updated Disinfection Kit User Manual. 2) If you have distributed affected devices, notify your customers and include a copy of this notice. 3) Please report all device-related incidents to the firm, distributor or local representative. 3) Complete and return the response form via email to [email protected] The Disinfection Kit User Manual updated to include: 7a: Perform Power-Out Alarm test Ensure steps 1-6 have been carried out, so that the alarm circuit is charged . Unplug the unit from the mains/utility power socket . Check that the audible alarm sounds for at least 120 seconds . The unit is now ready for storage . User Manuals can be downloaded from the following links: Airvo 2: https://resources.fphcare.com/content/airvo-manual-uk-us-and-az-ui-185045495-h-15thmarch22.pdf myAirvo 2: https://resources.fphcare.com/content/myairvo2-user-manual-ui-185045490.pdf Disinfection Kit: https://resources.fphcare.com/content/airvo-2-disinfection-manual-user-instructions-ui-185043723.pdf Please note, as per the "Technical Information" section of the Product Technical Manual and the User Manual, the expected service life is 5 years. If you have any questions, please contact the firm's Customer Care via email at [email protected] or directly at +1 (800) 446 3908 ext 5003 or +1 (949) 453 4000 ext 5003.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology