Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723 Recall
Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723
FDA recall Class II
Ongoing
FDA lists this Class II recall for Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0158-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 18, 2025
- FDA report date
- Oct 22, 2025
- Quantity in commerce
- 51,144
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723
Affected model & UDI checker
Search the exact identifiers published for Z-0158-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| PT101US | Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723 | Check source notice | Not specified |
| PT100US | Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723 | Check source notice | Not specified |
| UI-185043723 | Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723 | Check source notice | Not specified |
FDA code information
REF/UDI-DI: PT101US/9420012422347; PT100US/9420012422248. Disinfection Kit User Manual all revisions before UI-185043723 rev P
Why was this product recalled?
When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.
What actions did Fisher & Paykel Healthcare, Ltd. take?
On 9/18/2025, correction notices were communicated via email and phone call. Customers were asked to do the following: 1) Prior to use on a new patient, follow the instructions of the updated Disinfection Kit User Manual. 2) If you have distributed affected devices, notify your customers and include a copy of this notice. 3) Please report all device-related incidents to the firm, distributor or local representative. 3) Complete and return the response form via email to [email protected] The Disinfection Kit User Manual updated to include: 7a: Perform Power-Out Alarm test Ensure steps 1-6 have been carried out, so that the alarm circuit is charged . Unplug the unit from the mains/utility power socket . Check that the audible alarm sounds for at least 120 seconds . The unit is now ready for storage . User Manuals can be downloaded from the following links: Airvo 2: https://resources.fphcare.com/content/airvo-manual-uk-us-and-az-ui-185045495-h-15thmarch22.pdf myAirvo 2: https://resources.fphcare.com/content/myairvo2-user-manual-ui-185045490.pdf Disinfection Kit: https://resources.fphcare.com/content/airvo-2-disinfection-manual-user-instructions-ui-185043723.pdf Please note, as per the "Technical Information" section of the Product Technical Manual and the User Manual, the expected service life is 5 years. If you have any questions, please contact the firm's Customer Care via email at [email protected] or directly at +1 (800) 446 3908 ext 5003 or +1 (949) 453 4000 ext 5003.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0158-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026