DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513 Recall
DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
FDA recall Class II
Ongoing
FDA lists this Class II recall for DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0160-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 12, 2026
- FDA report date
- Oct 21, 2026
- Quantity in commerce
- 2,088
- Distribution
- See key details
- FDA cause
- Component design/selection
What is being recalled?
DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
Affected model & UDI checker
Search the exact identifiers published for Z-0160-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| C76947 | DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513 | Check source notice | Not specified |
| C49513 | DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513 | Check source notice | Not specified |
FDA code information
Product/UDI-DI: DxU 840m/15099590748654, DxU 850m/15099590748647. All Serial Numbers
Why was this product recalled?
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did BECKMAN COULTER, INC. take?
On 8/12/2026, recall notices were mailed and emailed to customers asking them to take the following actions: 1) Perform the daily Cleaning Instrument Surfaces maintenance described in Chapter 12, Maintenance Procedure of the DxU 840/850m Instructions for Use (PN# C49320AB), (PN# D38676AA), or iQ200 Instructions for Use (PN# 300-4320CE), (PN# 300-4321EE). During the cleaning, inspect the instrument for signs of lamina leakage, including fluid accumulation on or around the instrument, or evidence of fluid on the laboratory bench or floor. If you determine that there is a leak, discontinue use of the analyzer and clean any spilled fluid in accordance with your laboratory's safety procedure. 2) Contact the firm's Technical Support Organization immediately if evidence of leakage is observed. 3) Continue routine operation, if no leakage is observed, while maintaining increased awareness for the symptoms described in the impact section of this letter. 4) Post this letter on or near affected Analyzers. 5) Share this information with laboratory staff and retain this notification. If you have forwarded any of the affected products to another laboratory, please provide them a copy of this letter. Firm is investigating this issue and will continue to address any reported lamina fluid leakage through its Technical Support Organization until this issue has been fully resolved. If you have any questions regarding this notice, please contact the firm's Customer Support Center: http://www.beckmancoulter.com or 1-800-854-3633 in the United States and Canada.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0160-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology