FDA RECALL RECORD · Z-0166-2026

Cardiac Workstation 7000; Recall

Cardiac Workstation 7000; Model Number: 860441;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Cardiac Workstation 7000;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0166-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 12, 2025
FDA report date
Oct 22, 2025
Quantity in commerce
29 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Cardiac Workstation 7000; Model Number: 860441;

Affected model & UDI checker

Search the exact identifiers published for Z-0166-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
860441Cardiac Workstation 7000;Check source noticeNot specified
Not specifiedCardiac Workstation 7000;Check source notice00884838094833
FDA code information

Model Number: 860441; UDI-DI: 00884838094833; Serial Numbers: US42520284, US32520248, US32520244, US22520238, US22520221;

View the original FDA recall notice

Why was this product recalled?

Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.

What actions did Philips North America take?

On September 12, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Share this communication with all clinical staff to review and undedrstand. Please this letter with the device. Pass the notice along to all those who need to be aware within the organization or to any organization where affected devices were transferred. Ensure technical solution established by Philips is implemented as soon as available within the timeframe communicated by Philips. In case any device cannot be corrected, inform Philips of the reason why the Technical Solution could not be implemented. Action By Philips: A Philips representative will contact customers to arrange a replacement of the device's hardware on affected devices with the pinched power module wire within the housing of the Cardiac Workstation.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history