Brand Name: Branding is not applied to this product. Product Name: DVI display cables Recall
Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-12 Software Version: N/A Product Description: 12 foot DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with the 91390 and 91393
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Branding is not applied to this product. Product Name: DVI display cables. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0168-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 5, 2025
- FDA report date
- Oct 22, 2025
- Quantity in commerce
- 820
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-12 Software Version: N/A Product Description: 12 foot DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with the 91390 and 91393
Affected model & UDI checker
Search the exact identifiers published for Z-0168-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 012-0895-12 | Brand Name: Branding is not applied to this product. Product Name: DVI display cables | Check source notice | Not specified |
FDA code information
Lot Code: UDI-DI for products that are used with the display cables. (01)10841522131479 (Xprezzon), (01)10841522131455 (Qube)
Why was this product recalled?
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
What actions did Spacelabs Healthcare, Inc. take?
On 09/05/2025, the firm email an "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them that the DVI cables which are used for connecting the Qube (91390) or Xprezzon (91393) to the 94267-19, 94267-L19, 94267-L17 and 94267-L15 displays, do not meet the EMI/EMC electrical safety standards specifically for electromagnetic shielding and may pose serious risks to patient safety by interfering with the operation of another device, particularly in environments where electrically sensitive medical devices are in use. Customers are instructed to acknowledge receipt and understanding of this notification and request replacement cables as soon as possible, however no later than 31-Dec-2025. Do not replace cables while a patient is admitted to the monitor to mitigate any potential disruption in active patient monitoring. Any affected cables not currently in use should not be placed into service. For questions and/or assistance with this recall, contact: Spacelabs Technical Support Spacelabs Healthcare, Inc 35301 SE Center Street Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support Email: [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0168-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026