Medline surgical convenience kits labeled as: 1) NEURO-SPINE BASIN PACK, Medline Product Number/SKU DYNJ80055A; 2) AV SHUNT PACK-LF, Medline Product Number/SKU DYNJ24528M. Recall
Medline surgical convenience kits labeled as: 1) NEURO-SPINE BASIN PACK, Medline Product Number/SKU DYNJ80055A; 2) AV SHUNT PACK-LF, Medline Product Number/SKU DYNJ24528M.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Medline surgical convenience kits labeled as: 1) NEURO-SPINE BASIN PACK, Medline Product Number/SKU DYNJ80055A; 2) AV SHUNT PACK-LF, Medline Product Number/SKU DYNJ24528M.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0177-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 2, 2026
- FDA report date
- Oct 21, 2026
- Quantity in commerce
- 1900 kits
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Medline surgical convenience kits labeled as: 1) NEURO-SPINE BASIN PACK, Medline Product Number/SKU DYNJ80055A; 2) AV SHUNT PACK-LF, Medline Product Number/SKU DYNJ24528M.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 26BBC545; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 26ABM193; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 25LBD334; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 25JBL305; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 25IBF209; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 25HBI041; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 25FBF808; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 25EBO692; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 25DBG912; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 25BBS302; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 25ABN459; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 24LBQ417; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 24KBL387; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 24KBB074; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 24IBI827; Medline product number/SKU DYNJ80055A, UDI/DI 10198459093425 (each) 40198459093426 (case), Lot number 24HBL301; Medline product number/SKU DYNJ24528M, UDI/DI 10193489678222 (each) 40193489678223 (case), Lot number 22CLA280; Medline product number/SKU DYNJ24528M, UDI/DI 10193489678222 (each) 40193489678223 (case), Lot number 22ALA687; Medline product number/SKU DYNJ24528M, UDI/DI 10193489678222 (each) 40193489678223 (case), Lot number 21KLA284; Medline product number/SKU DYNJ24528M, UDI/DI 10193489678222 (each) 40193489678223 (case), Lot number 21JLA298; Medline product number/SKU DYNJ24528M, UDI/DI 10193489678222 (each) 40193489678223 (case), Lot number 21HLA613; Medline product number/SKU DYNJ24528M, UDI/DI 10193489678222 (each) 40193489678223 (case), Lot number 21GLA706; Medline product number/SKU DYNJ24528M, UDI/DI 10193489678222 (each) 40193489678223 (case), Lot number 21DLA076; Medline product number/SKU DYNJ24528M, UDI/DI 10193489678222 (each) 40193489678223 (case), Lot number 21ALA482; Medline product number/SKU DYNJ24528M, UDI/DI 10193489678222 (each) 40193489678223 (case), Lot number 20LLA600.
Why was this product recalled?
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
What actions did Medline Industries, LP take?
Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 9/2/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com, Recall Reference #: R-26-201-FG, Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable." Those consignees who further distributed the product were directed to notify the parties to whom the affected product was distributed. For questions, contact the Recall Department at 866-359-1704 or [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0177-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.