Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0182-2027

Convenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474F Recall

Convenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474F

FDA recall Class II

Ongoing

FDA lists this Class II recall for Convenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474F. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0182-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 31, 2026
FDA report date
Oct 21, 2026
Quantity in commerce
3,551 kits
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Convenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474F

Affected model & UDI checker

Search the exact identifiers published for Z-0182-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedConvenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474FCheck source notice10198459535772
Not specifiedConvenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474FCheck source notice10198459098512
Not specifiedConvenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474FCheck source notice10195327559007
FDA code information

DYNJG901004C UDI-DI 10198459535772 Lots 25LLA307 26BLA900 DYNJ88071 UDI-DI 10198459098512 Lots 24KMD416 24LMG268 25AMH987 25BMA076 25GMK411 DYNJ50474F UDI-DI 10195327559007 Lots 24DDA107 24DDA469 24EDA512 24GDB884 24IDA882

View the original FDA recall notice

Why was this product recalled?

Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medline Industries, LP take?

The firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology