Scout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use with the SCOUT MD Surgical Guidance System, Rx Only, Sterile-EO Recall
Scout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use with the SCOUT MD Surgical Guidance System, Rx Only, Sterile-EO
FDA recall Class II
Ongoing
FDA lists this Class II recall for Scout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use with the SCOUT MD Surgical Guidance System, Rx Only, Sterile-EO. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0187-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 27, 2026
- FDA report date
- Oct 21, 2026
- Quantity in commerce
- Not reported
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Scout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use with the SCOUT MD Surgical Guidance System, Rx Only, Sterile-EO
Affected model & UDI checker
Search the exact identifiers published for Z-0187-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Scout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use with the SCOUT MD Surgical Guidance System, Rx Only, Sterile-EO | Check source notice | 00884450392850 |
| Not specified | Scout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use with the SCOUT MD Surgical Guidance System, Rx Only, Sterile-EO | Check source notice | 00884450392874 |
| Not specified | Scout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use with the SCOUT MD Surgical Guidance System, Rx Only, Sterile-EO | Check source notice | 00884450392911 |
FDA code information
MDS05A-01-Lot: H3515477/UDI:00884450392850; MDS05C-01 - Lot: H3507786/UDI:00884450392874; MDS10C-01 - Lot:H3507794/UDI:00884450392911
Why was this product recalled?
The implantable clip may experience difficulties deploying, which could prevent the reflector from functioning as intended during use.
What actions did Merit Medical Systems, Inc. take?
On August 27, 2026, Merit Medical issued a Urgent Medical Device Recall Notice to affected consignees via FedEx. Merit ask consignees to take the following actions: 1. Immediately determine if any of the devices identified in the attached Customer ResponseForm (CRF) are within your facility. 2. Immediately discontinue the use of the affected lots of the SCOUT MD Delivery System. 3. Ensure that applicable personnel within your organization are made aware of this recall. 4. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. 5. Please fill out, scan and email the completed CRF to Customer Service at [email protected] within 7 business days. All affected products shipped to you must be accounted for on the CRF. 6. Immediately return all affected lots in your possession to Merit, per the instructions in the attached CRF.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0187-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data