Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0191-2027

Amplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3. 9-ACP2-007-020 4. 9-ACP2-007-022 5. 9-ACP2-010-025 6. 9-ACP2-010-028 7. 9-ACP2-010-031 8. 9-ACP2-010-034 Recall

Amplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3. 9-ACP2-007-020 4. 9-ACP2-007-022 5. 9-ACP2-010-025 6. 9-ACP2-010-028 7. 9-ACP2-010-031 8. 9-ACP2-010-034

FDA recall Class II

Ongoing

FDA lists this Class II recall for Amplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3. 9-ACP2-007-020 4. 9-ACP2-007-022 5. 9-ACP2-010-025 6. 9-ACP2-010-028 7. 9-ACP2-010-031 8. 9-ACP2-010-034. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0191-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 10, 2026
FDA report date
Oct 28, 2026
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
Labeling design

What is being recalled?

Amplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3. 9-ACP2-007-020 4. 9-ACP2-007-022 5. 9-ACP2-010-025 6. 9-ACP2-010-028 7. 9-ACP2-010-031 8. 9-ACP2-010-034

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI-DI 1. 00811806013459 2. 00811806013466 3. 00811806013473 4. 00811806013480 5. 00811806013497 6. 00811806013503 7. 00811806013510 8. 00811806013527 All lots/serial numbers.

View the original FDA recall notice

Why was this product recalled?

Changes to Instructions for Use (IFU) for the device regarding risk of Pulmonary Artery Injury (PAI).

What actions did Abbott take?

US IFU was updated and is available via https://manuals.eifu.abbott as of September 4, 2026. The firm began notifying consignees of the updated IFU on September 10, 2026. Abbott asks that physician users continue to follow all procedural steps that have already been provided via training, and that are now included in the IFU, as presented below: " After the landing zone has been identified, assess the closest distance to the pulmonary artery. " A distance =1.5 mm represents significant risk for a pulmonary artery injury; a distance between 1.5 and 3.0 mm represents a theoretical risk for a pulmonary artery injury; and a distance >3.0 mm represents a lower risk for a pulmonary artery injury. " To reduce the risk of pulmonary artery injury, consider an alternative landing zone or device orientation when clinically appropriate and C.L.O.S.E. criteria can still be met. " If the risk of pulmonary artery injury cannot be adequately minimized, the physician should carefully consider whether to proceed with device implantation.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology