Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0192-2027

Think Surgical Square Driver REF: 109126 Recall

Think Surgical Square Driver REF: 109126

FDA recall Class II

Ongoing

FDA lists this Class II recall for Think Surgical Square Driver REF: 109126. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0192-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 20, 2026
FDA report date
Oct 28, 2026
Quantity in commerce
15 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Think Surgical Square Driver REF: 109126

Affected model & UDI checker

Search the exact identifiers published for Z-0192-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
109126Think Surgical Square Driver REF: 109126Check source noticeNot specified
FDA code information

Lot: 25356/Primary DI: M9461091260

View the original FDA recall notice

Why was this product recalled?

Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did THINK Surgical, Inc. take?

On August 25, 2026, THINK Surgical Inc. issued a Urgent Medical Device Recall Notification to affected consignees via email. Think Surgical sent an updated letter via email on September 22, 2026 to provide additional product information. THINK Surgical ask consignees to take the following actions: 1. Examine your inventory of square drivers for the lot number 25356. Quarantined return any inventory of that lot to THINK Surgical, Inc., 2. If you have distributed or transferred affected devices to other facilities or end users, please promptly notify them and provide a copy of this notification., 3. Fill out the accompanying Acknowledgement Form. Email the completed Acknowledgement Form to THINK Surgical, Inc at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.