Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0194-2027

Medtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., Rxonly Recall

Medtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., Rxonly

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., Rxonly. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0194-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 31, 2026
FDA report date
Oct 28, 2026
Quantity in commerce
14210 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Medtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., Rxonly

Affected model & UDI checker

Search the exact identifiers published for Z-0194-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
8801071Medtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., RxonlyCheck source noticeNot specified
FDA code information

UDI: 20805832070016/ Lot codes: 2506331 2506571 2510111 2510511 2511091

View the original FDA recall notice

Why was this product recalled?

Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.

What actions did Medtronic Navigation, Inc. take?

On September 11, 2026 Medtronic Navigation Inc. issued a Urgent Medical Device Recall notification email. Medtronic ask consignees to take the following actions: 1. Immediately locate and quarantine all unused impacted product(s).; 2. Return unused impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form.; 3. Complete the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information, and email the completed form to Medtronic at [email protected].; 4. If the affected devices have already been utilized and/or discarded, we still ask that you complete and return the Customer Confirmation Form detailing that information.; 5. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records.; 6. If you purchase this product directly from NDI, please contact NDI for the list of impacted product and instructions for return: [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology