Philips Smart-hopping 2.0 AP 1.4 GHz; Product/ Recall

Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips Smart-hopping 2.0 AP 1.4 GHz; Product/. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0199-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 16, 2026
FDA report date
Oct 28, 2026
Quantity in commerce
174
Distribution
See key details
FDA cause
Software design

What is being recalled?

Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;

Affected model & UDI checker

Search the exact identifiers published for Z-0199-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
867216Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Check source noticeNot specified
FDA code information

Product/Model Number: 867216; UDI-DI: N/A; System Serial Numbers: DK61624330, DK61623830, DK61724824, DK61623802, DK61623826, DK61624101, DK61624082, DK61623910, DK61624187, DK51313245, DK61724472, DK61724620, DK61724590, DK61724461, DK61724725, DK61724679, DK61724684, DK51313645, DK51313455, DK61724814, DK61724747, DK61624235, DK61422548, DK52316132, DK61724553, DK61724957, DK61724756, DK61523178, DK61724502, DK61724860, DK61523117, DK61724577, DK61724376, DK61724566, DK61422553, DK61322150, DK61523070, DK61523270, DK52617072, DK61523151, DK61422454, DK53817745, DK52617073, DK61422674, DK61422973, DK61523211, DK61322124, DK61523272, DK61523248, DK51313320, DK61422463, DK61523260, DK61322137, DK61322337, DK61322253, DK61322375, DK61322380, DK61422762, DK61422868, DK61322122, DK61422851, DK61322283, DK61523658, DK61422848, DK61422884, DK61422897, DK61523409, DK61422808, DK61422736, DK61422806, DK61422763, DK61523517, DK61422957, DK61523383, DK61523442, DK61322227, DK61422843, DK61523495, DK61423030, DK61422949, DK61322271, DK61322127, DK60621383, DK60621482, DK61322213, DK60321063, DK60320788, DK60621426, DK60621646, DK61322238, DK60621223, DK60621517, DK61322379, DK52316002, DK60320966, DK60621224, DK60621197, DK60821808, DK60621565, DK60821916, DK60821936, DK60320570, DK60320720, DK60320768, DK60320506, DK60320537, DK60321057, DK60320711, DK60320892, DK60320761, DK60320544, DK60321033, DK60320571, DK60320759, DK60320484, DK60321044, DK60320839, DK60320419, DK60320437, DK60320474, DK54820263, DK60320601, DK60320592, DK54820262, DK54820037, DK54819845, DK54820134, DK54819871, DK54820272, DK60320874, DK54519084, DK54519572, DK54820343, DK54819876, DK52516932, DK54519153, DK53817766, DK54518977, DK52516923, DK54519018, DK53817821, DK54820004, DK54819819, DK54820101, DK54819922, DK54518948, DK52516510, DK54519031, DK54518951, DK54519115, DK52516754, DK52516873, DK52516766, DK53817862, DK52516569, DK52516783, DK52516933, DK54418622, DK52516533, DK52315958, DK53217276, DK52316143, DK53217187, DK53217154, DK52316114, DK53217197, DK54418453, DK53918024, DK54418698, DK54418600, DK54418588, DK54418881, DK54519407, DK54418598;

View the original FDA recall notice

Why was this product recalled?

Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).

What actions did Philips North America take?

On september 16, 2026, URGENT Medical Device Correction, letters were sent to customers. Actions to be taken: 1. Direct clinical attendance of the patient during transport, when this issue is most likely to manifest, should be considered if not currently a part of the clinical workflow. In the event a persistent No Central Monit technical alarm is noted at the wireless device or a No Data technical alarm is noted at the central monitor; a restart of the wireless monitor may restore central monitoring when the patient is in known good wireless coverage area. 2. Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. Actions planned by Philips to correct the problem 3. Philips representative will contact you to arrange the software upgrade SH 2.0 AP 1.4GHz which will correct this issue. If you need any further information, please contact your local Philips representative, or call (800) 722-9377.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.