Product: STA Liatest D-Di; REF: 00515; Recall
Product: STA Liatest D-Di; REF: 00515;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Product: STA Liatest D-Di; REF: 00515;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0205-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 12, 2025
- FDA report date
- Oct 22, 2025
- Quantity in commerce
- 12740 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Product: STA Liatest D-Di; REF: 00515;
Affected model & UDI checker
Search the exact identifiers published for Z-0205-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 00515 | Product: STA Liatest D-Di; REF: 00515; | Check source notice | Not specified |
| Not specified | Product: STA Liatest D-Di; REF: 00515; | Check source notice | 03607450005158 |
FDA code information
REF: 00515; UDI-DI: 03607450005158; Lot numbers: 271382, 271411; Expiration 09/30/2025;
Why was this product recalled?
A potential risk of underestimation of D-Dimer (D-Di) levels with the two listed lots.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Diagnostica Stago, Inc. take?
On September 12, 2025 URGENT: MEDICAL DEVICE RECALL (REMOVAL) letters were emailed to customers. Actions for customers: Discontinue use of affected lots and dispose of them in accordance with your local regulations. Complete and return response form. For additional information, please contact Stago's 24/7 Hotline at 1-800-725-0607.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0205-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology