Techno-Path Multichem AE Unassayed Level 3, REF:AE623X Recall
Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
FDA recall Class II
Ongoing
FDA lists this Class II recall for Techno-Path Multichem AE Unassayed Level 3, REF:AE623X. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0207-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 14, 2026
- FDA report date
- Oct 28, 2026
- Quantity in commerce
- 832 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
Affected model & UDI checker
Search the exact identifiers published for Z-0207-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| AE623X | Techno-Path Multichem AE Unassayed Level 3, REF:AE623X | Check source notice | Not specified |
FDA code information
UDI:(01)05391523442824/-Lot # 65705263
Why was this product recalled?
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
What actions did TECHNO-PATH MANUFACTURING LTD. take?
On August 14, 2026, Techno Path Manufacturing Ltd has issued a medical device correction to consignees via email. Techno Path Manufacturing Ltd informed consignees to please verify their inventory against the affected product codes and lot numbers. Techno Path Manufacturing Ltd. has informed the consignees to take the following actions: 1. Immediately place all affected inventory on hold and prevent further distribution. 2. Quarantine all affected product currently under your control. 3. Confirm that there has been no distribution of the impacted lots to customers. 4. Confirm receipt of this notification and provide inventory and distribution details to Technopath Manufacturing Ltd. within 3 business days. 5. Arrange return or destruction of affected inventory in accordance with instructions provided by Technopath Manufacturing Ltd.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0207-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.