FDA RECALL RECORD · Z-0241-2026

Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system Recall

Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system

FDA recall Class II

Ongoing

FDA lists this Class II recall for Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0241-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 29, 2025
FDA report date
Oct 29, 2025
Quantity in commerce
29 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system

Affected model & UDI checker

Search the exact identifiers published for Z-0241-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedZenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray systemCheck source notice00884838091535
FDA code information

Product Number (REF): 718096; UDI-DI: 00884838091535; All serial numbers;

View the original FDA recall notice

Why was this product recalled?

Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?

On September 29, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Circulate this notice to all users and follow the instructions in the letter. 2. Affected systems may continue to be used in accordance with their intended use and Instructions for Use (IFU). 3. Ensure the letter is in a place likely to be seen/viewed. 4. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions by Philips: Starting in October 2025, Philips will replace the 11W control board of the affected Zenition 50 systems. Your local Philips representative will contact you to schedule a visit to perform this activity in your system.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.