Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system Recall
Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
FDA recall Class II
Ongoing
FDA lists this Class II recall for Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0241-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 29, 2025
- FDA report date
- Oct 29, 2025
- Quantity in commerce
- 29 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
Affected model & UDI checker
Search the exact identifiers published for Z-0241-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system | Check source notice | 00884838091535 |
FDA code information
Product Number (REF): 718096; UDI-DI: 00884838091535; All serial numbers;
Why was this product recalled?
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?
On September 29, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Circulate this notice to all users and follow the instructions in the letter. 2. Affected systems may continue to be used in accordance with their intended use and Instructions for Use (IFU). 3. Ensure the letter is in a place likely to be seen/viewed. 4. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions by Philips: Starting in October 2025, Philips will replace the 11W control board of the affected Zenition 50 systems. Your local Philips representative will contact you to schedule a visit to perform this activity in your system.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0241-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology