Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0250-2027

Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620 Recall

Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620

FDA recall Class II

Ongoing

FDA lists this Class II recall for Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0250-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 16, 2026
FDA report date
Oct 28, 2026
Quantity in commerce
598
Distribution
See key details
FDA cause
Process control

What is being recalled?

Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620

Affected model & UDI checker

Search the exact identifiers published for Z-0250-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedDirexion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620Check source notice08714729839613
FDA code information

Material Number: M001195620; UDI-DI: 08714729839613; Batch Numbers: 37775591, 37886793, 37914472, 37914473, 37957744, 38112479, 38149826, 38167704, 38175797, 38444076, 38446234, 38530457, 38745190, 38990885, 39140998, 39140999, 40470358, 40499682, 40499683, 40502254, 40504732, 40504733, 40507436, 40524004, 40539147, 40539467;

View the original FDA recall notice

Why was this product recalled?

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

What actions did Boston Scientific Cork, Ltd. take?

On September 16, 2026, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: 1. Immediately stop further use or distribution and segregate affected product (see Attachment 1 for identification of the affected devices). 2. Forward this notice to relevant personnel and any facilities to which the product was transferred. 3. Continue to follow the standard of practice when monitoring patients treated with the product. 4. Complete and return the enclosed Reply Verification Tracking Form per the included instructions. 5. Return product to Boston Scientific in accordance with the enclosed instructions.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology