Halyard ANTERIOR HIP PACK. Recall
Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH30.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Halyard ANTERIOR HIP PACK.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0283-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 8, 2026
- FDA report date
- Oct 28, 2026
- Quantity in commerce
- 80 kits
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH30.
Affected model & UDI checker
Search the exact identifiers published for Z-0283-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| SOAH30 | Halyard ANTERIOR HIP PACK. | Check source notice | Not specified |
FDA code information
Model/Catalog Number: SOAH30. UDI-DI: 10809160520246. Lot Code: Lot Numbers: 1705826, 1705827. Expiration Date: 1/31/2028.
Why was this product recalled?
A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.
What actions did AVID Medical, Inc. take?
AVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to identify, segregate, and quarantine all impacted kits identified, and to not use the impacted component located in the piggybacked portion of the convenience kit. Customers were also instructed to add warning labels to affected kits to ensure all users are aware of the actions to remove and discard the impacted component prior to use. distributors were instructed to notify consignees if affected units were further distributed. Additionally, consignees were requested to complete and return the Recall Response Form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0283-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology