Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging Recall

Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

FDA recall Class II

Ongoing

FDA lists this Class II recall for Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0291-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 12, 2025
FDA report date
Oct 29, 2025
Quantity in commerce
64 units
Distribution
See key details
FDA cause
Package design/selection

What is being recalled?

Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Affected model & UDI checker

Search the exact identifiers published for Z-0291-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
4XXElscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imagingCheck source noticeNot specified
FDA code information

all serial numbers in distribution

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 9/12/2025 via letter (USPS first class). The notice explained the issue, potential safety issue, and requested the following: "Actions to be taken by Customer/User: You have previously been formally notified that these Impacted Systems have reached their End of Guaranteed Service life. GE HealthCare does not have the ability to correct these devices. GE HealthCare requests you to immediately STOP USING your Impacted System(s). As these devices are no longer supported and are no longer serviceable, continued use may compromise patient or user safety, negatively impact product performance, or introduce other risks to patients, users, and facilities. This is the final communication from GE HealthCare regarding these devices. Please ensure all potential users in your facility are made aware of this Stop Use safety notification. Take care when deinstalling the system, ensuring the system is balanced and all parts are secure. If you require assistance in removal and disposal of your system, please contact your local GE HealthCare representative. If you would like to discuss your replacement options, please contact your GE HealthCare account representative. Please retain this document for your records. Contact Information: If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative." On 10/01/2025, the firm sent a follow-up notice removing Millenium VG from this action as it was erroneously included in the first notice.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.