Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0291-2027

ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V Recall

ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V

FDA recall Class II

Ongoing

FDA lists this Class II recall for ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0291-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 12, 2026
FDA report date
Nov 4, 2026
Quantity in commerce
31 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V

Affected model & UDI checker

Search the exact identifiers published for Z-0291-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
INFX-8000VALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000VCheck source noticeNot specified
FDA code information

DI-04987670100222/

View the original FDA recall notice

Why was this product recalled?

Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.

What actions did Canon Medical System, USA, INC. take?

On October 12, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction to affected consignees. Canon provided the following to consignees: 1. A representative of Canon Medical Systems USA will contact you to schedule a time to conduct installation of modified software to prevent internal errors from occuring if F-STORE or image acquisition is performed while CAAS clinical analysis function is used. 2. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at (800) 521-1968 or by email [email protected]. 3. In order to comply with the FDA record keeping requirement, Canon respectfully requests that the attached form be completed and returned via email to [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology