FDA RECALL RECORD · Z-0292-2027

Philips Azurion 3 M12. Model Numbers: 722063, 722221. Recall

Philips Azurion 3 M12. Model Numbers: 722063, 722221.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips Azurion 3 M12. Model Numbers: 722063, 722221.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0292-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 6, 2026
FDA report date
Nov 4, 2026
Quantity in commerce
95 units
Distribution
See key details
FDA cause
Equipment maintenance

What is being recalled?

Philips Azurion 3 M12. Model Numbers: 722063, 722221.

Affected model & UDI checker

Search the exact identifiers published for Z-0292-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
722063Philips Azurion 3 M12. Model Numbers: 722063, 722221.Check source noticeNot specified
722221Philips Azurion 3 M12. Model Numbers: 722063, 722221.Check source noticeNot specified
FDA code information

1. Model Number: 722063. UDI Numbers: (01)00884838085275(21)56, (01)00884838085275(21)96, (01)00884838085275(21)98, (01)00884838085275(21)77, (01)00884838085275(21)145, (01)00884838085275(21)47, (01)00884838085275(21)80, (01)00884838085275(21)146, (01)00884838085275(21)117, (01)00884838085275(21)99, (01)00884838085275(21)104, (01)00884838085275(21)3, (01)00884838085275(21)144, (01)00884838085275(21)62, (01)00884838085275(21)61, (01)00884838085275(21)22, (01)00884838085275(21)153, (01)00884838085275(21)40, (01)00884838085275(21)70, (01)00884838085275(21)30, (01)00884838085275(21)25, (01)00884838085275(21)121, (01)00884838085275(21)18, (01)00884838085275(21)14, (01)00884838085275(21)16, (01)00884838085275(21)17, (01)00884838085275(21)53, (01)00884838085275(21)101, (01)00884838085275(21)143, (01)00884838085275(21)41, (01)00884838085275(21)151, (01)00884838085275(21)93, (01)00884838085275(21)150, (01)00884838085275(21)105, (01)00884838085275(21)129, (01)00884838085275(21)21, (01)00884838085275(21)78, (01)00884838085275(21)23, (01)00884838085275(21)92, (01)00884838085275(21)59, (01)00884838085275(21)139, (01)00884838085275(21)154, (01)00884838085275(21)75, (01)00884838085275(21)27, (01)00884838085275(21)24, (01)00884838085275(21)94, (01)00884838085275(21)100, (01)00884838085275(21)26, (01)00884838085275(21)106, (01)00884838085275(21)37, (01)00884838085275(21)36, (01)00884838085275(21)135, (01)00884838085275(21)123, (01)00884838085275(21)89, (01)00884838085275(21)86, (01)00884838085275(21)79, (01)00884838085275(21)133, (01)00884838085275(21)81, (01)00884838085275(21)140, (01)00884838085275(21)52, (01)00884838085275(21)48, (01)00884838085275(21)148, (01)00884838085275(21)55, (01)00884838085275(21)113. Serial Numbers: 56, 96, 98, 77, 145, 47, 80, 146, 117, 99, 104, 3, 144, 62, 61, 22, 153, 40, 70, 30, 25, 121, 18, 14, 16, 17, 53, 101, 143, 41, 151, 93, 150, 105, 129, 21, 78, 23, 92, 59, 139, 154, 75, 27, 24, 94, 100, 26, 106, 37, 36, 135, 123, 89, 86, 79, 133, 81, 140, 52, 48, 148, 55, 113. 2. Model Number: 722221. UDI Numbers: (01)00884838099203(21)6, (01)00884838099203(21)93, (01)00884838099203(21)115, (01)00884838099203(21)177, (01)00884838099203(21)124, (01)00884838099203(21)106, (01)00884838099203(21)151, (01)00884838099203(21)71, (01)00884838099203(21)159, (01)00884838099203(21)60, (01)00884838099203(21)33, (01)00884838099203(21)55, (01)00884838099203(21)44, (01)00884838099203(21)142, (01)00884838099203(21)144, (01)00884838099203(21)126, (01)00884838099203(21)147, (01)00884838099203(21)26, (01)00884838099203(21)23, (01)00884838099203(21)133, (01)00884838099203(21)167, (01)00884838099203(21)15, (01)00884838099203(21)74, (01)00884838099203(21)202, (01)00884838099203(21)22, (01)00884838099203(21)58, (01)00884838099203(21)165, (01)00884838099203(21)78, (01)00884838099203(21)174, (01)00884838099203(21)176, (01)00884838099203(21)207. Serial Numbers: 6, 93, 115, 177, 124, 106, 151, 71, 159, 60, 33, 55, 44, 142, 144, 126, 147, 26, 23, 133, 167, 15, 74, 202, 22, 58, 165, 78, 174, 176, 207.

View the original FDA recall notice

Why was this product recalled?

Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.

What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?

Philips Medical notified consignees on about 10/06/2026 via certified mailed letter titled "URGENT Medical Device Correction." Consignees were instructed that affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the additional instructions provided, which include: if motorized movements become unavailable, the stand can be manually repositioned using the handgrips and brake controls located on both sides of the stand. For ceiling-mounted systems, manual longitudinal (1), transversal (2), and rotational (3) stand movements remain available. Note: Transversal stand movements (2) are only available for ceiling-mounted systems with the FlexMove option. For floor-mounted systems, only rotational (3) stand movements remain available. The table longitudinal position (4) can be adjusted manually when the table is not tilted. Additionally, consignees were instructed to ensure the notification letter is distributed to all relevant users, maintained with the device documentation until corrective actions are completed, to post the letter near affected devices for ease of reference, notify customers/organizations if affected systems have been distributed or transferred, and complete and return the response form provided. Philips will inspect all systems in scope and replace SSDs that are not specified in the Medical Device File. Philips expects to start the implementation of this correction in December 2026. A local Philips representative will contact consignees to schedule a visit for a Field Service Engineer to perform these activities.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.