Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0312-2026

Brand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MM Recall

Brand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MM Model/Catalog Number: EK087P Software Version: N/A Product Description: REDUCING CONVERTER 10/12MM TO 5MM Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MM. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0312-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 24, 2025
FDA report date
Nov 5, 2025
Quantity in commerce
155 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MM Model/Catalog Number: EK087P Software Version: N/A Product Description: REDUCING CONVERTER 10/12MM TO 5MM Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-0312-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
EK087PBrand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MMCheck source noticeNot specified
Not specifiedBrand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MMCheck source notice04046963620059
FDA code information

Model No EK087P; UDI-DI: 04046963620059; Lots 52974276, 52952133, 52954633, 52959357, 52964442, 52968887, 52972214, 52977427, 52981208, 52985519, 52988828, 52992514, 52993617, 52997919, 53002094, 53004347, 53006380, 53007625, 53010869, 53013557

View the original FDA recall notice

Why was this product recalled?

It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did AESCULAP, INC. take?

On September 24, 2025 URGENT MEDICAL DEVICE RECALL NOTIFICATION letters were sent to customers. Actions to be taken. 1. Review the recall notice and ensure all users in the organization are notified. Forward this notification to customers. Remove affected units from inventory, quarantine the product but do not destroy it. Complete and return the included response form. Return affected units to the firm. Credit of the product will be issued. An alternative disposable universal seal is available and can be ordered.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.