FDA lists this Class II recall for SuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0319-2026
- Recall classification
- Class II
- Enforcement status
- Terminated
- Initiation date
- Sep 16, 2025
- FDA report date
- Nov 5, 2025
- Quantity in commerce
- 45 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
SuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system
Affected model & UDI checker
Search the exact identifiers published for Z-0319-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | SuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system | Check source notice | 00863157000336 |
FDA code information
UDI/DI 00863157000336
Why was this product recalled?
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
What actions did Surgical Theater Inc take?
Surgical Theater issued an URGENT: Medical Device Software correction notice dated September 17, 2025 to its consignees on 09/22/2025 via email. The notice explained the issue, risk to health, and provided the following: Customers can continue to use the software version(s) noted above until a software patch is made available by Surgical Theater. Until the software patch is installed, please follow these steps: 1. Read and share this notice with all relevant staff. 2. Continue to use the 3D Fusion functions to review the fusion results for all case files to verify alignment of secondary layer(s) to primary layer and to adjust fusion as necessary, and 3. Acknowledge your receipt of this notification by completing and returning the included acknowledgement form via email to [email protected]. A software patch is currently undergoing verification and should be available in the next 10 days. Surgical Theater has revised the 2D Image Fusion function to ensure the x, y and z axis rotation values are consistently applied when generating the aligned Axial, Sagittal and Coronal images. In addition, the rotate left/right buttons in all three planes were modified so the volume rotation stays consistent with the respective plane orientation. A Surgical Theater representative will contact your facility to schedule installation once the patch is available. For questions regarding this notification contact: Kevin Murrock, Sr. Director of Quality & Regulatory, Tel. 330-472-6520 (Monday through Friday, 9:00 AM to 5:00 PM, Eastern Time), email: [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Recall termination recorded
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0319-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology