Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0327-2026

VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Recall

VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

FDA recall Class II

Ongoing

FDA lists this Class II recall for VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0327-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 16, 2025
FDA report date
Nov 5, 2025
Quantity in commerce
2 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

Affected model & UDI checker

Search the exact identifiers published for Z-0327-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MRT-3020/U7VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7.Check source noticeNot specified
MEXL-3020/B7VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7.Check source noticeNot specified
FDA code information

CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.

View the original FDA recall notice

Why was this product recalled?

Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.

What actions did Canon Medical System, USA, INC. take?

Cannon Medical began notifying consignees via email on about 10/16/2025. Consignees were instructed that a representative of Canon Medical Systems USA will contact them to schedule a time to replace the relevant parts. If the message "Gradient power supply error is displayed before the replacement is performed, immediately stop using the system and contact your service representative. Consignees were requested to complete and return the provided response form

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology