BioZorb Marker. Radiographic soft tissue marker. Recall
BioZorb Marker. Radiographic soft tissue marker.
FDA recall Class I
Ongoing
FDA lists this Class I recall for BioZorb Marker. Radiographic soft tissue marker.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0328-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Oct 10, 2024
- FDA report date
- Dec 4, 2024
- Quantity in commerce
- 12,710 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
BioZorb Marker. Radiographic soft tissue marker.
Affected model & UDI checker
Search the exact identifiers published for Z-0328-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| F0405 | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | Not specified |
| Not specified | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | 15420045514065 |
| Not specified | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | 15420045514058 |
| Not specified | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | 15420045514041 |
| Not specified | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | 15420045514034 |
| Not specified | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | 15420045514027 |
| Not specified | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | 15420045514010 |
| Not specified | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | 15420045514003 |
| Not specified | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | 15420045513990 |
| Not specified | BioZorb Marker. Radiographic soft tissue marker. | Check source notice | 15420045513983 |
FDA code information
Model No. F0405, F0404, F0331, F0231, F0221, F0304, F0303, F0203, F0202; UDI-DI: 15420045514065, 15420045514058, 15420045514041, 15420045514034, 15420045514027, 15420045514010, 15420045514003, 15420045513990, 15420045513983; Lot No. 22E24RH, 23A09RK, 23L23RP, 22E24RG, 23A09RJ, 23E03R, 23G28RW, 22F06RL, 22G26RQ, 22K25RA, 23A09RH, 23A30RP, 23C02RJ, 23H25RW, 23K04RC, 23K31RG, 24A30RG, 22D21RBB, 22F06RK, 22F29RK, 22G26RP, 22H24RM, 22J26RG, 22K25R, 22L22R, 22M20RM, 23A26RV, 23C02RH, 23E27RF, 23F29RT, 23H03R, 23H25RU, 23K04RA, 23L23RK, 24B02RD, 24B27RK, 22D21RAA, 22E24RA, 22F29RJ, 22G26RN, 22H24RL, 22J26RF, 22K21R, 22L22RA, 22M20RL, 23A26RU, 23C02RG, 23C30RQ, 23E03RA, 23F29RP, 23G28RV, 23H25RV, 23K02RB, 23K31RD, 23L23RH, 24A08RY, 24B01R, 24B27RL, 24C22RN, 22D21RZ, 22E24RB, 22F25RF, 22G26RM, 22H24RK, 22J26RD, 22K19RW, 22L22RB, 23C02RK, 23E25RY, 23F29RS, 23H24RP, 23K04RB, 23K31RF, 23L23RJ, 22D21RY, 22E24R, 22F25RD, 22G26RL, 22H22RT, 22J26RC, 22K19RV, 22L22RC, 23A09RN, 23H03RA, 23H31R, 23K04RD, 23K25RM, 22D21RX, 22E20RAA, 22F21R, 22G21RM, 22H22RS, 22J22RBB, 22K19RU, 22L22RF, 23A09RM, 23A30RQ, 23B24RZ, 23C30RP, 23D27RJ, 23E31RD, 23F28RZ, 23G31RJ, 23H24RQ, 23K02RA, 23K25RN, 22D20RV, 22E20RZ, 22F16RV, 22G21RL, 22H20R, 22J22RAA, 22K19RT, 22K25RB, 22L22RD, 23A09RL, 23B24RY, 23C28RJ, 23D27RH, 23E30RF, 23F29RQ, 23G31RH, 23H24RN, 23K02R, 23K25RL.
Why was this product recalled?
Firm received reports of adverse events/complications that include serious injuries, such as pain, infection, rash, device migration, device erosion, seroma, and discomfort.
What actions did Hologic, Inc. take?
Consignees were sent an URGENT: MEDICAL DEVICE RECALL letter by mail dated 10/24/24. The notification instructs consignees to quarantine all affected devices and share the recall notification with all surgeons in their practice that have used or planned to use the BioZorb Marker. Additionally, if product was further distributed the notice is to be forwarded to those customers. Affected devices will be returned to Hologic and coordination of return will be initiated by return of the Customer Response Form. Healthcare providers are recommended to be aware of reports of serious adverse events following the placement of the BioZorb Marker devices in breast tissue and to continue to monitor patients for signs of any adverse events. BioZorb Markers are not cleared to fill space in the tissue or improve cosmetic outcomes after procedures. Patients are recommended to contact their healthcare provider if they experience any adverse events following the placement of their BioZorb marker. Hologic states that there is no need to undergo explantation of the device unless advised by your healthcare provider. Patients who have received the BioZorb Marker and/or your provider is planning radiation therapy treatments, discuss the possible risks with your provider. Report any problems/complications and direct any questions regarding this recall to Hologic at [email protected]. ***Update May 5, 2026: On April 6, 2026 A Follow-up notification as an updated resource for clinicians who have previously used Biozorb breast markers or are currently caring for patients implanted with BioZorb markers. Hologic recalled and removed from market BioZorb and BioZorb LP on October 24, 2024; therefore, all patients with these implants have had them in place for at least one year. The original notification that was distributed as a part of the previous recall can be found as an appendix to this resource.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0328-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology