Halyard MINOR PACK. Recall
Halyard MINOR PACK. Model/Catalog Number: VASF105-08.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Halyard MINOR PACK.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0331-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 8, 2026
- FDA report date
- Nov 11, 2026
- Quantity in commerce
- 108 kits
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Halyard MINOR PACK. Model/Catalog Number: VASF105-08.
Affected model & UDI checker
Search the exact identifiers published for Z-0331-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| VASF105-08 | Halyard MINOR PACK. | Check source notice | Not specified |
FDA code information
Model/Catalog Number: VASF105-08. UDI-DI: 10809160451960. Lot Number: 1659384. Expiration Date: 9/30/2027.
Why was this product recalled?
Potential for open seals on header bags, compromising sterility.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did AVID Medical, Inc. take?
AVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to Segregate, Quarantine, and Inspect all kits identified. If the header bag seal appears to be compromised, discard the entire kit. Consignees were also requested to complete and return the provided Recall Response Form. Distributors were instructed to notify customers if affected kits were further distributed. On 10/06/2026 AVID Medical sent an updated "URGENT: MEDICAL DEVICE CORRECTION" letter to consignees via email. The strategy was updated and consignees were instructed not to use the affected kits and that they should either discard or return them. If returning, they should contact customer service for a Return Goods Authorization. Consignees were again requested to complete and return the provided Recall Response Form. Distributors were again instructed to notify customers if affected kits were further distributed.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0331-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology