Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Recall
Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0349-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 30, 2026
- FDA report date
- Nov 11, 2026
- Quantity in commerce
- 4 device in the U.S.
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.
Affected model & UDI checker
Search the exact identifiers published for Z-0349-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| CI1000 | Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 | Check source notice | Not specified |
FDA code information
Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286
Why was this product recalled?
Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits
What actions did Cochlear Limited take?
On 09/30/2026, an "URGENT Field Safety Corrective Action Notification" was sent to U.S. customers informing them that a potential condition has been identified that could result in one or more open circuits affecting individual electrode pathways with the implant. Customers were instucted to not implant and return affected products to Cochlear Limited. For questions contact [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0349-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology