FDA RECALL RECORD · Z-0353-2027

Operating Table TruSystem 7000 U (dV) Recall

Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633

FDA recall Class II

Ongoing

FDA lists this Class II recall for Operating Table TruSystem 7000 U (dV). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0353-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 29, 2026
FDA report date
Nov 11, 2026
Quantity in commerce
62 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633

Affected model & UDI checker

Search the exact identifiers published for Z-0353-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1723633Operating Table TruSystem 7000 U (dV)Check source noticeNot specified
FDA code information

Serial numbers: 108920104, 110192607, 110192885, 110208936, 110212885, 110220066, 110227520, 110248337, 110250614, 109423464, 110192608, 110192886, 110208941, 110212886, 110220068, 110227523, 110248343, 110250621, 109517138, 110192609, 110206656, 110210717, 110220055, 110220069, 110227524, 110249706, 110250622, 110139767, 110192683, 110206657, 110210718, 110220056, 110220071, 110227525, 110249712, 110206662, 110139776, 110192882, 110206659, 110210720, 110220057, 110220072, 110227552, 110250605, 110188921, 110188919, 110192883, 110206660, 110210721, 110220058, 110220078, 110227553, 110250609, 110188922, 110188920, 110192884, 110206664, 110212881, 110220065, 110220079, 110248218, 110250613; UDI-DI number: 00887761967519

View the original FDA recall notice

Why was this product recalled?

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Baxter Healthcare Corporation take?

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 9/29/26 was sent to customers. Actions to be Taken by Customers 1. A Baxter representative will contact your facility when the correction for this issue becomes available. Baxter expects to begin contacting customers with more information about the correction by the end of 2026. 2. Customers should continue to monitor the table as instructed in the applicable IFU during use for unusual noises or changes in performance. If an unusual noise is observed, discontinue use of the affected product and contact Baxter Technical Service for assistance. 3. Please post this communication by the impacted devices until the correction is completed to maintain user awareness. 4. If you received this communication directly from Baxter, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com. Log in to the portal using the account number listed in the enclosed reply instruction sheet. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification. 5. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 6. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 7. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box o

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology