Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Recall
Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0366-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 22, 2025
- FDA report date
- Nov 5, 2025
- Quantity in commerce
- 3318 units (169 US, 3149 OUS)
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6;
Affected model & UDI checker
Search the exact identifiers published for Z-0366-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| L2KPP2 | Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; | Check source notice | Not specified |
| L2KPP6 | Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; | Check source notice | Not specified |
| Not specified | Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; | Check source notice | 00630414961910 |
| Not specified | Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; | Check source notice | 00630414961927 |
FDA code information
1. Siemens Material Number (SMN): 10381441; Catalog Number: L2KPP2; UDI-DI: 00630414961910; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389; 2. Siemens Material Number (SMN): 10381442; Catalog Number: L2KPP6; UDI-DI: 00630414961927; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389;
Why was this product recalled?
The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Siemens Healthcare Diagnostics, Inc. take?
On September 22, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Discontinue use of and discard the kit lot listed in the table above (Products Section). Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. For continued Intact PTH testing in your laboratory during this time, please contact your Siemens representative to discuss alternative Siemens Healthineers solutions. As an alternative, customers who have an ADVIA Centaur system or an Atellica IM/CI may utilize the corresponding iPTH assay on those systems. These platforms are not impacted. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. A follow-up communication will be provided when assay performance has been restored. ***Update 2/18/2026*** On February 10, 2026 Follow-up Urgent Medical Device Correction letters were sent to additional customers to notify of the addition of new lots for the recall. Customer actions remained the same as the original letter.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0366-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology