Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322 Recall

Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322

FDA recall Class II

Ongoing

FDA lists this Class II recall for Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0408-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 8, 2025
FDA report date
Nov 5, 2025
Quantity in commerce
504,000
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322

Affected model & UDI checker

Search the exact identifiers published for Z-0408-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
454322Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322Check source noticeNot specified
Not specifiedGreiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322Check source notice29120017574488
FDA code information

UDI-DI: 29120017574488(Rack), 39120017574485(Case). Lot: B250139Q, Expiration: 01/09/2026

View the original FDA recall notice

Why was this product recalled?

Blood collection tubes were manufactured with the label applied too high on the tube, which shifted the printed fill level mark upward, which could result in unnecessary sample rejection and repeat collections.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Greiner Bio-One North America, Inc. take?

Greiner issued an Urgent Product Recall notice to its consignees on 08/19/2025 via email and fax. The notice explained the issue and requested the following: "Immediate Actions Required 1. Stop Distribution & Isolate Inventory Immediately cease distribution of the affected lot and segregate all remaining units. 2. Select ONE of the following notification options: A. Customer Notification by Greiner Bio-One o Complete the attached Product Disposition Site Confirmation form. o Email the form to [email protected] or fax to 877-217-7927. o Provide an Excel list of all customers including, if applicable, subdistributors who received the affected product, containing: . Facility name and address (city, state, ZIP) . Contact name, email, and phone number for Clinical/Laboratory Director or Supply Chain . Date(s) and quantity shipped o Greiner will handle direct customer notification and coordinate product return/credit. B. Customer Notification by Distributor o Use the customer letter provided by Greiner Bio-One. o Contact all customers who received the affected lot, including any sub-distributors. o Conduct follow-up with non-responders and document effectiveness checks per FDA guidance: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts/industryguidance-recalls" For questions regarding this recall, please contact: Greiner Bio-One Technical Service, Phone: 800-515-8112, Fax: 877-217-7927, Email: [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology