Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Recall
Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: Ultrasound device for intraoperative use Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0410-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 26, 2025
- FDA report date
- Nov 5, 2025
- Quantity in commerce
- 14
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: Ultrasound device for intraoperative use Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-0410-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MQC04001 | Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System | Check source notice | Not specified |
| MQC14001 | Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System | Check source notice | Not specified |
| MQU14011 | Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System | Check source notice | Not specified |
| MQV12011 | Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System | Check source notice | Not specified |
| 5042 | Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System | Check source notice | Not specified |
FDA code information
Lot Code: Serial number; Model number 5042; MQC04001 5043; MQC04001 5044; MQC14001 5061; MQU14011 5062; MQU14011 5064; MQU14011 5065; MQU14011 5066; MQU14011 5067; MQC04001 5037; MQU14011 5073; MQV12011 5038; MQU14011 5045; MQU14011 5063; MQU14011
Why was this product recalled?
Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.
What actions did Medistim Asa take?
Firm notified impacted consignees by email on September 26, 2025. Consignees were informed that qualified service personnel will perform the correction of the device according to the standard service procedure.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0410-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology